BNC-2115 Installation Guide
8
ni.com
To connect DIO signals to the NI PCI-60
XX
E and NI AT-MIO-64E-3
devices, use the AI <16..55> BNC-2115 connectors. To connect DIO
signals to the NI PCI-6025E and NI AT-MIO-16DE-10 devices, use the
P0.<0..7>, P1.<0..7>, and P2.<0..7> BNC-2115 connectors.
For more detailed information about how the accessory terminals
correspond one-to-one to the Multifunction DAQ device, refer to the I/O
Connector Pinouts Table in the
E Series Help
at
ni.com/manuals
.
Operating the BNC-2115 in Single-Ended Mode
You can increase the number of analog input channels by configuring
the DAQ device analog input to operate in either referenced single-ended
(RSE) mode, or nonreferenced single-ended (NRSE) mode. In either of
these modes, each BNC coaxial cable carries two single-ended channels,
instead of one differential channel.
To configure the analog input using Traditional NI-DAQ, open
Measurement & Automation Explorer (MAX), open
Devices and
Interfaces»Traditional NI-DAQ Devices
, then right-click the icon for
your device. Select
Properties
from the menu, then click the
AI
tab and
change the analog input (AI) mode to RSE or NRSE.
To configure input terminals using NI-DAQmx, use the DAQ Assistant or
the NI-DAQmx API. For more information, refer to the
DAQ Assistant Help
and the
Measurement and Automation Explorer Help for NI-DAQmx
.
Table 4.
BNC-2115 Two-Position Terminal Block Connections for
Extended DIO Accessory
Two-Position
Terminal Block
Connector
Signal Description
D GND
Digital Ground—This pin supplies the
reference for the digital signals at the I/O
connector as well as the +5 VDC supply.
P1.3
Port 1 line 3—This is the bidirectional
DIO line for Port 1.