National Instruments Corporation
17
NI 9235/9236 Calibration Procedure
Specifications
The values in the following table are based on calibrated scaling
coefficients, which are stored in the onboard EEPROM. The following
calibration specifications are for 23 ±5 °C.
Table 2.
NI 9235/9236 Measurement Accuracy
Module
Percent of Reading
(Gain Error)
Percent of Range*
(Offset Error)
NI 9235/9236
0.04% max
0.002% max
* Range equals 29.4 mV/V.
Table 3.
NI 9235/9236 Shunt Calibration Accuracy
Module
Percent of Reading (Gain Error)
NI 9235
0.22% max
NI 9236
0.20% max
Table 4.
NI 9235/9236 Excitation Voltage Accuracy
Module
Accuracy
NI 9235/9236
±1% max