National Instruments Corporation
11
NI 9235/9236 Calibration Procedure
10. Acquire 5,000 points of data using the DAQmx Read VI, as shown in
the following table.
11. Average the readings that you acquired and record the result.
12. Clear the acquisition using the DAQmx Clear Task VI, as shown in the
following table.
13. Repeat steps 4 through 12 for all calibrator output values.
NI recommends that you verify all values, although you can save time
by verifying only the values used in your application.
14. Perform the following calculation for each test result value other than
0 V/V, using the results you recorded in step 11.
Test Result
–
0 V/V Test Result
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxReadAnalogF64
with
the following parameters:
taskHandle
:
ta
s
kHandle
numSampsPerChan
:
–1
timeout
:
10.0
fillMode
:
DAQmx_Val_GroupByChannel
readArray
:
data
arraySizeInSamples
:
5000
sampsPerChanRead
:
&read
reserved
:
NULL
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxClearTa
s
k
with the
following parameter:
taskHandle
:
ta
s
kHandle