NI 4065 Calibration Procedure
84
ni.com
Completing the Adjustment Procedures
Call the niDMM Close External Cal VI to complete the adjustment
procedure for the NI 4065 and close the session.
Verification Limits
This section includes the verification limits for DC voltage, AC voltage,
4-wire resistance, 2-wire resistance, DC current, and AC current for the
NI 4065. Compare these limits to the results in the
section.
Note
Use the values in the
24-Hour Limits
column for a post-adjustment verification
only
.
Otherwise, use the values in the
1-Year Limits
column.
Limits in the following tables are based upon the August 2010 edition of the
NI 4065
Specifications
. Refer to the most recent NI 4065 Specifications online at
ni.com/
manuals
. If a more recent edition of the specifications is available, recalculate the limits
based upon the latest specifications.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CloseExtCal
with the
following parameters:
Instrument_Handle
:
Cal Session
Action
:
NIDMM_EXTCAL_ACTION_SAVE
if the
results of the calibration are satisfactory,
and you want to save the new calibration
coefficients to the EEPROM.
Otherwise,
Action
:
NIDMM_EXTCAL_ACTION_ABORT
if the
results of the calibration are
unsatisfactory, and you want to restore
the original calibration coefficients to the
EEPROM.