12 Specifications
124
NBM-550
Narda
12.1 Display
12.2 Measurement functions (with HF probes)
Display type
Transreflective monochrome LCD
Display size
10 cm (4"), 240 x 320 pixels
Backlight
White LEDs, selectable lighting time (OFF, 5 s,
10 s, 30 s, 60 s, PERMANENT)
Display refresh rate
200 ms for bar graph and graphics,
400 ms for numerical result values
Result units
mW/cm
2
, W/m
2
, V/m, A/m, % of standard
(% of standard is related to power density)
Display range
0001 to 9999, switchable between variable and permanent triads
Display range,
variable triads
0.01 V/m to 100.0 kV/m
0.01 mA/m to 265.3 A/m
0.001 mW/m
2
to 26.53 MW/m
2
0.1 nW/cm
2
to 2.653 kW/cm
2
0.0001% to 9999%
Display range,
fixed triads
0.01 to 9999 V/m
0.0001 to 265.3 A/m
0.0001 to 9999 W/m
2
0.0001 to 9999 mW/cm
2
0.0001% to 9999%
Result types
(isotropic, RSS)
Latest value (Actual), Maximum, Minimum,
Average, Maximum Average
Result type (XYZ mode)
Latest values: Actual X, Actual Y, Actual Z
(for probes with separate axis connections)
Time averaging
Averaging time selectable from 4 s to 30 min (2 s steps)
Spatial averaging
Individual or continuous sampling
Multi-position spatial
averaging
Averaging over up to 24 spatially averaged results,
storage of individual position results
and of the total value
Progress memory
(History mode)
Graphical display of the latest values versus time
(time span 2 min to 8 h)
Correction frequency
1 kHz to 100 GHz or OFF (direct frequency selection,
interpolation between calibration points)
Hot Spot search
Acoustic indication of increasing or decreasing
field strength (Actual or Maximum)
Alarm function
2 kHz warning tone (4 Hz repetition rate), variable threshold
Timer controlled recording
(Timer Logging)
Starting time setting: Up to 24 h in advance or record immediately
Recording time: up to 100 h
Store interval: Every 1 s to 6 min (in 11 steps)
Number of store intervals: up to 32000
Содержание NBM-550
Страница 10: ...Contents VIII NBM 550 Narda ...
Страница 38: ...3 Preparing the NBM 550 for use 28 NBM 550 Narda ...
Страница 52: ...4 Getting started 42 NBM 550 Narda ...
Страница 88: ...6 Measurements with the NBM 550 and EHP 50 78 NBM 550 Narda ...
Страница 110: ...9 PC software 100 NBM 550 Narda ...
Страница 139: ...12 10 CE Declaration of Conformity Narda NBM 550 129 12 10 CE Declaration of Conformity ...
Страница 140: ...12 Specifications 130 NBM 550 Narda ...
Страница 141: ...12 11 Declarations of origin Narda NBM 550 131 12 11 Declarations of origin NBM 550 Germany EHP 50 Italy ...
Страница 142: ...12 Specifications 132 NBM 550 Narda ...
Страница 146: ...13 Ordering information 136 NBM 550 Narda ...
Страница 147: ...Narda NBM 550 137 14 Glossary This chapter explains some important terms that are used in this operating manual ...
Страница 154: ...Index 144 NBM 550 Narda ...
Страница 155: ......