6.10 The Marker function
Narda
IDA
-
3106
109
6.10.2 Using the markers separately
You can set and display the two markers A and B independently using the
Marker: Single
marker function.
To use the markers separately:
You have selected the necessary trace types (see
The
Marker
sub-menu is active.
1.
Press the
Marker
softkey until
Marker: Single
is shown.
2.
Press the
Active Marker
softkey to select marker
A
or
B
.
3.
Select a trace type / detector: Press the
Active Trace
softkey, select an
entry with the rotary control knob and then press the
OK
key.
You can now use the available functions separately for markers A and B:
NOTE:
An algorithm that covers all the existing measurement values contained in
the spectrum is used for peak markers. Peak values are indicated by “Pk”
after the trace type / detector (e.g.: Max Pk: 105.8 mV/m). If the marker is
not on a peak value, “Val” is shown after the trace type / detector (e.g.: Max
Val: 105.8 mV/m). Another special feature of the peak marker is that the
frequency is tracked and follows the peak value (Peak Tracking: On).
Table 28: Marker functions for selecting peak values
Selection
Function
Rotary control knob
Moves the marker to the desired position manually.
Highest Peak
softkey
Places the marker on the highest peak value.
Softkey
Next Peak
Places the marker on the next higher peak value.
Peak Left
softkey
Places the marker on the next peak value to the left of
the previous position.
Peak Right
softkey
Places the marker on the next peak value to the right
of the previous position.
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......