6.6 Setting the reference level (Reference Level)
Narda
IDA
-
3106
99
6.6.2
Automatic reference level search (RL Search)
This softkey initiates a search for a suitable reference level. This is
recommended if the field situation is unknown.
Press the
RL Search
softkey.
The message:
Searching for best reference level. Please wait
. is
displayed. A measurement over the entire frequency range covered
by the antenna you are using is made in the background. This takes
several seconds. When it is completed, the device sets the reference
level to the optimum value automatically and reverts to normal
measurement mode.
NOTE:
Press the
ESC
key if you want to cancel the process.
RL Search Type
The IDA
-
3106 provides two types of reference level search which can be
selected using the
RL Search Type
softkey (toggle function):
•
Sensitive:
This type is suitable for all largely static signals.
•
Safe:
This type allows for a higher overload reserve. It is better to set the
reference level manually if pulsed signals are being measured.
To select the search type:
Press the
RL Search Type
softkey.
The switch toggles between the two types each time the softkey is
pressed.
6.6.3
Measurements with preamplifier
You can activate a 20 dB preamplifier in the antenna handle when
measuring very small signals. The reference level correction is
automatically taken into account.
To activate / deactivate the preamplifier:
A Narda antenna handle is connected.
1.
Press the
Reference Level
softkey, then press the
Preamp:
softkey.
The softkey indicates the staus of the preamplifier.
2.
Press
ESC
to return to the measurement menu.
Содержание IDA
Страница 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Страница 8: ...Contents VI IDA 3106 Narda...
Страница 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Страница 134: ...6 Common Functions 126 IDA 3106 Narda...
Страница 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Страница 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Страница 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Страница 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Страница 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Страница 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Страница 246: ...15 PC Software 238 IDA 3106 Narda...
Страница 254: ...16 Remote Control 246 IDA 3106 Narda...
Страница 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Страница 272: ...17 Specifications 264 IDA 3106 Narda...
Страница 279: ...Narda IDA 3106 271 19 Glossary The meanings of terms and abbreviations are explained in this chapter...
Страница 284: ...19 Glossary 276 IDA 3106 Narda...
Страница 290: ...Index 282 IDA 3106 Narda...
Страница 291: ......