ELT-400
5 Operation and use
5-9
5.3.3
Selecting the detection mode (DETECT)
Three modes of detection are available depending on the selected operating
mode:
ª
Detection mode / Detector: see page C-2
⇒
Press the
DETECT
key until the required detection mode is shown in the
display.
Note:
The selected detection mode is applied immediately.
The measurement value is valid as soon as the switch over occurs.
Detection mode
Field
strength
mode
Exposure
STD mode
STND (Standard)
The detection mode specified in the appropriate
standard is set automatically. You can also select RMS
and PEAK detection if required.
x
RMS (root mean square value)
The RMS value resulting from the isotropic
measurement is derived from the RMS values for the
three individual measurement axes.
x
x
PEAK (peak value)
The peak value of the isotropic measurement is
derived from the peak value of the field vector. The
phase in the three individual measurement axes is
taken into account.
x
x
Table 5-2
Available detection modes, depending on the selected operating mode
Содержание ELT-400
Страница 2: ......
Страница 3: ...ELT 400 Exposure Level Tester BN 2304 xx Series A Operating Manual...
Страница 8: ...Notes...
Страница 12: ...1 Introduction ELT 400 1 4 Notes...
Страница 16: ...2 Safety instructions ELT 400 2 4 Notes...
Страница 22: ...3 Preparation for use ELT 400 3 6 Notes...
Страница 44: ...5 Operation and use ELT 400 5 16 Notes...
Страница 68: ...7 Remote control ELT 400 7 22 Notes...
Страница 95: ...ELT 400 B field probe connector RS 232 Scope Charge...
Страница 96: ...Notes...
Страница 100: ...Notes...