Cat.No.
B03-02-J
Model configuration
On-site type pressure transmitter for process use.
Since it is for general applications and proof
pressure explosion-protected use, select the type
corresponding to the working environment.
•
Since the sensor uses a semiconductor strain
gauge, durability and stability are excellent.
•
Sinse the wetted parts use a high
corrosion-resistant metal diaphragm, corrosion
resistance is excellent and use with a wide range
of gases and liquids is possible.
•
EMI countermeasures have been taken so that
transceivers and other electronic devices are not
affected.
KH41
KH43
KD41, KD43
General process use
Explosion-proof use (Explosion-proof construction)
KH41
KD41 (d2G4)
KH43
KD43 (d2G4)
Fluid:
Gas, liquid
*The diaphragm type is available, when the viscosity of
fluid to be measured is high. Please contact us.
Operating condition:
KH41, 43 (General process use)
Under the normal condition,where there is no
inflammable gas or liquid which cause the ignition or
explosion.
KD41, 43 (Explosion-proof construction)
Hazardous area
For details, please refer to the description of the
proof pressure explosion-protected construction.
Type:
Direct mounting, surface mounting, pipe mounting
Connection:
Rc 1/2, Rc 1/4, G3/8B (Diaphragm only)
Wetted parts material:
KH41 SCS14 + SUS316
KH43 Co-Ni alloy
KD41 SCS14 + SUS304
KD43 SCS14 + Co-Ni alloy
Pressure range:
-5 to +5→-50 to +50kPa
-0.1 to 0→-0.1 to 2MPa
0 to 5kPa→0 to 50MPa
Power source:
12 to 32V DC (Standard 24V DC)
Output:
4 to 20mA DC (2 wire system)
Load resistance:
600Ω Maximum (24V DC)
Withstand voltage:
250V AC, 1 minutes
Withstand voltage:
250V AC, 1 minutes
Allowable maximum pressure:
150% of pressure range
Operating temperature range:
KH41, 43 -30 to 80℃ (No freezing or condensation)
KD41, 43 -20 to 60℃ (No freezing or condensation)
Storage temperature range:
KH41, 43 -30 to 80℃ (No freezing or condensation)
KD41, 43 -30 to 70℃ (No freezing or condensation)
Operating humidity range:
30 to 95%RH (No condensation)
Accuracy:
±0.25%F.S. or ±0.5%F.S.
(Includes Linearity, hysteresis, repeatability)
Temperature coefficient:
±0.025%F.S./℃ (Zero)
±0.025%F.S./℃ (Span)
(-20 to 70℃)
Outlet for electric wire:
KH41, 43 PF1/2
KD41, 43 Proofpressure gasket type, conduit type
(Adjustable cable diameter φ7 to φ12)
Case material:
ADC12
Protection:
K□41 Splash-proof (IP54)
K□43 Jet-proof (IP55)
However, splash-proof in case of the pressure
range 2MPa or lower (IP54)
Weight:
Approx. 0.85kg to 5.8kg
KH41
・
43, KD41
・
43
Process Applications
Pressure Transmitter
Features
Outline
Specifications