9
Y-KH15-007A Rev.3
*1 Do not seal the cable end with taping or adhesive during wiring as it will block atmospheric pressure introduced via the cable.
Avoid using a cable with poor ventilation (no clearance within the cable).
*2 Max. allowable pressure is the upper limit of pressure value which may safely be applied to the product and remain in
specification once pressure is returned to the rated range with a couple of times overpressurization for about 10 minutes. Effects
of continuous overpressure are not guaranteed.
*3 Accuracy of +/-0.25%F.S. cannot be guaranteed for the model KH15-
□□
6 and 0 to 5V DC output.
*4 It will be 0 to 70°C for the pressure range of 0.1 MPa or less. For the wetted part exceeding a temperature of 70°C such as
steam or hot water, be sure to install a pipe siphon or similar device to keep it below 70°C. Also, when measuring
low-temperature fluid such as cold water, please employ preventive measures against condensation.
*5 IP67 waterproof case does not cover the waterproofing of the wiring opening (cable gland (JIS F 8801)).
*6 Indoor power distribution network to connect the product shall not be affected by lightning surge voltage and the switching
transients of the power system.
*7 Pressure range of 120 MPa is guaranteed only when Co-Ni alloy materials and the G1/2B fitting are combined.
5. DIMENSIONS
■
5.1 KH15-J
□□
(small terminal box)
< Figure 5-1 >
22 X 25.4 HEX.
(Unit: mm)