CO2 SHAKING INCUBATOR
MODEL : NB-206CXL/206CXXL
OPERATING MANUAL
Страница 1: ...CO2 SHAKING INCUBATOR MODEL NB 206CXL 206CXXL OPERATING MANUAL...
Страница 2: ...ecaution for use 6 2 4 Precaution for ground connection 7 3 FEATURE AND SPECIFICATION 3 1 Features 8 3 2 Specification 9 3 3 Configuration 10 11 4 INSTALLATION AND OPERATION 4 1 Control Panel of Incub...
Страница 3: ...ree warranty service will be provided once the unit is proved to be defective by wrong workmanship after NBIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of inst...
Страница 4: ...roduct and the plug must be 20cm at least The power outlet must be only for this product Using various products simultaneously can cause a fire Clean the power plug with a dry towel and connect it pro...
Страница 5: ...ndoor temperature is 20 30 Do not put inflammable substances near the product It can cause a fire CAUTION The product requires the distance of at least 20 cm from the wall for well ventilation When ve...
Страница 6: ...an cause a fire and an explosion Do not disassemble the shaker during the product is operating CAUTION Do not wash the product with excessive quantity of water thinner benzene and Petroleum It can cau...
Страница 7: ...ment to prevent elect leakage An electric shock an electric leakage and a fire can be occurred without an electric leakage breaker Do not ground to these places Gas Pipe water pipe pipe lighting rod t...
Страница 8: ...lass inner doors allow minimizing the loss of inner temperature CO2 density humidity when opening door UV LAMP It is located above ceiling in chamber and decontaminates air circulating from the fan Th...
Страница 9: ...r volume 650 liter 850 liter Shelves 3ea Chamber dimensions 700 W x 650 D x 1430 H mm 700 W x 800 D x 1530 H mm Each compartment dimensions Compartment 1 Bottom 700 W x650 D x430 H mm Compartment 2 3...
Страница 10: ...0 10 24 3 3 CONFIGURATION REAR VIEW RPM RS232 PORT is optional with additional charge SLINDING SHAKER PLATFORM Fuse Overheating Protection Thermal Cut Off Power Outlet CO Gas Inlet CONTROL PANEL GLASS...
Страница 11: ...TFORM Lock UV LAMP UV lamp near circulation humidity fan works for sterilization even during cell culturing It is covered by the fan sealing of chamber so it doesn t have a danger of contamination and...
Страница 12: ...SIGNAL Over Heating Protection Activated Alarm B HEAT OPERATING SIGNAL When Heater make heat red light on C CO SIGNAL When CO valve open and inject CO2 D ALARM S W Audible Alarm ON OFF E ALARM MUTE T...
Страница 13: ...tor Clear the air passage for gas input gasket at the rear of the unit Also check the gas tube and get rid of any obstacles for smooth gas flow Before supply of CO2 gas to incubator by gas tube check...
Страница 14: ...wer S W Then the LED screen will display the temperature in the chamber b Press the TEMPSET key Then the LED screen will flicker and display c Then input the desired temperature by pushing UP and DOWN...
Страница 15: ...n No DISPLAY FUNCTION 1 Chamber Temperature 2 Out Door Heater Temperature 3 Glass Door Heater Temperature 4 CO Calibration 5 Heating control 6 CO gas supply control 7 To apply a new value a Press and...
Страница 16: ...k CAL SET Door Frame Heater calibration Channel 3 is purposed to remove excessive water condensing on glass door caused by high temperature difference between chamber and outside When failed to remove...
Страница 17: ...rer Push UP and DOWN to set the value NOTE Channel 6 is to set starting point of solenoid control for CO2 supply Factory pre programmed set point is at 1 2 for NB203XL 3 for NB203 to optimize set up a...
Страница 18: ...the product at about 42 You can identify its activation point by lighting of pilot lamp At the set point turn the safety dial and recognize its deviation Considering the deviation set the safety 5 hig...
Страница 19: ...PM value b Press the SET button then RPM display will flicker c You can set the RPM through pressing keys during the display is flickering d After setting the desired value press the SET key again the...
Страница 20: ...he minute through the same way with setting the hour d Press the SET key after setting the time hour and minute then the screen displays t SAVE as below Time set point range is maximum 47hr 59min When...
Страница 21: ...PM In this case remove some flasks or lower RPM until it works well 4 Sliding Platform Push the red button at each side of front and slide it out to front Carefully slid it out to avoid flasks fallen...
Страница 22: ...elves are all demountable Followings are the step to disassemble shakers and shelves Ensure the product is turned off completely Unplug the shaker cable connector at the back of the chamber Photo 1 Un...
Страница 23: ...s alarm starts ringing So temp alarm tolerance delay is about 8 minutes When you press alarm mute during alarm ringing alarm immediately stops and 10 minutes alarm delay will be given If temp deviatio...
Страница 24: ...www n biotek com N BIOTEK OM NB 206CXL NB 206CXXL Ver 2020 24 24 7 APPREANCE DRAWING...