 Incubator Operation Manual 

 

 

 
 
 
 
 
 
 
 

Model: NB-203 / NB-203XL 

 

 
 
 
 

 

 

 

 

 

 

 

 

 
 
 
 
 
                                                   
 
 
 
 
 
 

 
 

 
 

Содержание NB-203

Страница 1: ...1 CO2 Incubator Operation Manual Model NB 203 NB 203XL...

Страница 2: ...PORTATION STORAGE AND LOCATION OF INSTALLATION 5 6 3 PREREQUISITE AND CONFIGURATION 7 8 4 FEATURES AND SPECIFICATIONS OF THE UNIT 9 5 CONTROL PANEL 10 6 OPERATION 11 14 7 OVERHEATING SAFETY 15 8 SERVI...

Страница 3: ...g the power cable Do not make the power plug be pressed by back of the product A space between the product and the plug must be 30cm at least The power outlet must be only for this product Using vario...

Страница 4: ...have the outlet for AC 220V then bury it under the ground after connecting the ground line to copper plate No ground connection can bring an electrocution an electric leakage and a Fire 1 4 Precaution...

Страница 5: ...of the door The inflow of open air can cause the change of temperature in chamber and discoloration of the packing part by a foreign substance 2 TRANSPORTATION STORAGE AND LOCATION OF INSTALLATION 2...

Страница 6: ...y accessible and can be easily pulled in case of danger Install the unit at a flat surface free from vibration and in a well ventilated location If the ground is not flat it can cause an excessive vib...

Страница 7: ...able head 5 The place where is being vibrated 6 Too narrow to use lift handle at side of bottom and power cable of incubator Cleaning before use Before conducting cell culture it is recommended to cle...

Страница 8: ...t CO2 leakage Outer door Water pan Heated by 6 sides direct heating for humidity Outer door heater prevent condensation CO2 inlet Safety switch Power cord Fuse holder NOTE 1 The figure 1 is to show yo...

Страница 9: ...Room Temperature 23 to 60 Accuracy 0 25 at 37 0 5 at 37 Controller Microprocessor Digital PID Humidity 70 80 at 37 CO2 Range 0 to 20 Accuracy 0 1 at 5 at 37 Increment 0 1 Sensor IR CO2 Sensor Outer do...

Страница 10: ...4 Temperature Display 5 CO Display 6 CO2 valve Control Indicator Flashing if solenoid valve of CO2 gas control is open 7 Alarm ON OFF Switch If turned off alarm is disarmed and no warning with audible...

Страница 11: ...ll Set up Temperature range is Ambient 5 60 Normal In case of low temperature control option like NB 203C NB203XLC turn on Cooling S W when low temperature setup is required The maximum controllable l...

Страница 12: ...perature 3 Glass Door Heater Temperature 4 CO Deviation Calibration 5 Heating On Off Range Control Factory Setup Only DO NOT CHANGE 6 CO gas supply speed control Factory Setup Only DO NOT CHANGE 7 To...

Страница 13: ...libration Channel 3 is purposed to remove excessive water condensing on glass door caused by high temperature difference between chamber and outside When failed to remove water condensing by calibrati...

Страница 14: ...From 3 of CO to setting value solenoid valve takes control of CO supply until it gets to setting value When above stages are cleared please press the CAL SET button to save the new value 6 3 ALARM 3...

Страница 15: ...personnel When replace any electrical or mechanical components the disconnect units should be far away from its electrical power source A Electrical connection Try to disconnect main power cable and...

Страница 16: ...s again there is short circuit contact N BIOTEK service Switch defective Replace the switch Gas CO2 concentration in chamber is too high too low Defective function of the CO2 controller Reset the alar...

Страница 17: ...rvice will be provided once the unit is proved to be defective by wrong workmanship after N BIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of installation or 1...