a. Press and hold
“CAL/SET” for 10 seconds. Then, LED will be flickering as below.
Channel 1 is at chamber
’s Main Temp calibration stage.
Press UP (▲) as much as difference from measured value by precise analyzer if it is higher.
Press DOWN (▼) as much as difference from measured value by precise analyzer if it is lower.
Ex) If measured temperature is 38
℃
and Display shows 37
℃
, then press up 1
℃
.
Note
* Calibration range for temperature is
±
5
℃
* To go to next channel is to press
“CAL/SET” button. After 5
th
channel, the LED is back to
temperature display.
No Function! Just skip this stage by pressing set button once.
No Function! Just skip this stage by pressing set button once.
c. Calibration Setting: Heating control
NOTE
Channel 5 is to set heating control point and is designed to prevent a significant temperature overshoot. This
mode is preprogrammed in the factory and should NOT be adjusted by the user.