N-BIOTEK NB-101S Скачать руководство пользователя страница 1

 

 

 

 
 
 
 
 
 
 
 
 

ORBITAL SHAKER 

 

 

MODEL : NB-101S

 

 
 
 

 

 
 
 

                                     

 
 
 
 
 

 
 
 
 

   

 

 

 
 

OPERATING MANUAL

 

Содержание NB-101S

Страница 1: ...ORBITAL SHAKER MODEL NB 101S OPERATING MANUAL...

Страница 2: ...have to use the product under specified electrical conditions If you use the product at the place in different power pressure it causes a fire and malfunction of the product Do not make the power plug...

Страница 3: ...k and a fire 1 2 Precaution for installation Use it with the proper voltage Please check the voltage hertz written on serial label Over voltage low voltage can damage the product and poor performance...

Страница 4: ...ult than the original purpose Do not use the product for different purpose It can cause malfunction or poor function Consequently you will get a wrong result Do not use an inflammable spray near the p...

Страница 5: ...an electrocution when malfunction or an electric leakage occurs At the place where you can t ground Please buy the equipment to prevent any electrical leakage An electric shock an electric leakage an...

Страница 6: ...uction drive Brush less DC motor provides less vibration and maintenance Artificial intelligence system which maintains precise speed adjust time LED indicates temperature speed time and state of powe...

Страница 7: ...tion Power switch PLATFORM CONTROL PANEL 2 4 Control panel A B C D E A DISPLAY For displaying RPM and TIME B RPM TIME Key for selecting RPM and TIME C SET Key for saving desired RPM and TIME D UP DOWN...

Страница 8: ...RPM setting range is from 30 to 300 If you do not press the SET key to store the desired speed the new speed will not be saved and the set speed will return to the prior setting 3 Time Set up 3 1 Pres...

Страница 9: ...ntinually you must set value as 00 00 When you set the time like above and operate instrument END is displayed after a minutes But shaking is well running without stop till reaching desired time After...

Страница 10: ...atform 101S SR Spring rack 101S H10 100ml Flask Holder x 16ea with 101S PL 101S H25 250ml Flask Holder x 9ea with 101S PL 101S H50 500ml Flask Holder x 5ea with 101S PL 101S H100 1000ml Flask Holder x...

Страница 11: ...r cable and then reconnect 2 Voltage supply Please check to voltage 3 Fuse check 4 Verify voltage on unit 5 Defective power switch check voltage at switch 2 This message may be displayed as to RPM is...

Страница 12: ...Mode 5 This message may be displayed as to The electricity is on from power failure during TIME Momentary noise symptom When the platform is being moved faster than RPM which 5 This message may be ma...

Страница 13: ...ree warranty service for the faulty caused by user s improper operation excessive use use of incorrect voltage frequency storage in wrong environment mentioned in Manual 4 Complete the above table aft...

Отзывы: