OPERATOR'S MANUAL
65
12.2. TECHNICAL MAINTENANCE
These instructions describe the maintenance procedures for the extraoral X-ray system. These instructions apply to all the revisions of said equipment.
In order to ensure the operational safety and functional reliability of the equipment installed, at least once a year an authorized service technician must perform a full inspection of the device.
When taking measurements that require a multimeter, always use a calibrated digital multimeter.
All the following tests will be carried out. Customer should be notified prior to replacing parts.
Answer questions with yes (√) or no (−)
Inspection DATE
Step
Description
Reference in User Manual
__/__/20__
__/__/20__
__/__/20__
1
Check that all labels located
-
on side of the device
-
at the base of the column,
-
on the X-ray tube housing,
are intact, well attached and legible.
Section Identification label position
2
Check there are no external damages to the equipment which could reduce protection against radiation.
Section Description of the operation
3
Check the remote emission button cable is not broken or scratched.
Section X-ray emission remote control
4
Check the power switch verifying that the switch is working properly and the green light on the switch, check if the keyboard
illuminates when the switch is in the on position.
Section Turning on the system
5
Verify that exposure is immediately interrupted when X-rays button is released
Section X-ray emission remote control
6
Check the control keyboard functionality: functions must respond to interrogation.
Section Console onboard the machine?
7
Check proper functioning of X-ray exposure LED and exposure buzzer
Section X-ray emission remote control
8
Check proper functioning of patient centering laser traces
Section Laser positioning system
9
Perform a motors movement test running the Dummy panoramic procedure, making sure the axis movement (rotation and
translation simultaneously) is smoothly and without noise. See figure 1, next page.
Section Perform a simulation (Dummy Run)
10
Perform column movements, up and down on the Z axis and check proper functioning. See figure 2, next page.
Section Console onboard the machine?
11
Check the emergency button functionality. Emergency button is used to stop X-ray device operation, it’s located near main switch
button.
Section Emergency stop button
12
Check the X-Ray Generator functionality performing a complete trial exam exposure. Select any panoramic exam and reset the
unit. Hold down the emission button throughout the entire exam procedure. Absence of error messages assures a proper
generator functionality.
Section Performing a 2D X-Ray examination
13
Perform a complete 2D calibration of the unit : column calibration, PAN x-ray alignment, PAN detector calibration, PAN mechanical
alignment, Laser test.
Technical manual, section 2D Calibration
14
At the end make a 2D calibration backup
Technical manual, section Calibration backup
15
(
Only for 3D machines
) Perform a complete 3D calibration of the unit: Beam limiter test, Daily check, Phantom acquisition cylindrical test. Technical manual, section 3D Calibration
16
(
Only for 3D machines
) At the end make a 3D calibration backup
Technical manual, section 3D Calibration backup
17
(
Only for 3D machines
) Perform an exam on the 3D Quality Phantom and evaluate the quality of the outcome volumetric study.
Acquisition Operations with Annex
Operator Name
Signature
The undersigned confirms that he/she has checked the unit for the above criteria and that he/she has handed over the unit in fully functional condition.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!!
Содержание 708G
Страница 1: ...97050799 Rev 05 16 09 EN IT FR DE ES PT RU TR ZH NO SV FI DA PL CS HU RO UK ...
Страница 2: ...2 OPERATOR S MANUAL EN ITALIANO ...
Страница 67: ...66 OPERATOR S MANUAL EN Figure 1 Figure 2 ...
Страница 132: ...66 ISTRUZIONI PER L USO IT Figura 1 Figura 2 ...
Страница 197: ...66 INSTRUCTIONS D UTILISATION FR Figure 1 Figure 2 ...
Страница 263: ...DE GEBRAUCHSANLEITUNG 67 Abbildung 1 Abbildung 2 ...
Страница 328: ...66 INSTRUCCIONES DE USO ES Figura 1 Figura 2 ...
Страница 393: ...66 INSTRUÇÕES PARA O USO PT Figura 1 Figura 2 ...
Страница 458: ...66 ИНСТРУКЦИИ ПО ПРИМЕНЕНИЮ RU Рисунок 1 Рисунок 2 ...
Страница 523: ...66 KULLANIM TALİMATLARI TR Şekil 1 Şekil 2 ...
Страница 574: ...52 操作人员手册 ZH 9 3 二维检查模式的等计量曲线没有相关图像说明 9 4 CBCT 检查的等剂量线 只限于 3D 版本的机器 ...
Страница 588: ...66 操作人员手册 ZH 图 1 图 2 ...
Страница 653: ...66 BRUKSANVISNING NO Figur 1 Figur 2 ...
Страница 718: ...66 INSTRUKTIONER FÖR ANVÄNDNING SV Figur 1 Figur 2 ...
Страница 769: ...52 KÄYTTÖOHJEET FI 9 3 ISODOOSIT 2D TUTKIMUKSILLE 9 4 ISODOOSIKÄYRÄT CBCT TUTKIMUKSILLE Vain 3D laitteelle ...
Страница 783: ...66 KÄYTTÖOHJEET FI Kuva 1 Kuva 2 ...
Страница 848: ...66 BRUGSANVISNING DA Figur 1 Figur 2 ...
Страница 913: ...66 INSTRUKCJA OBSŁUGI PL Rysunek 1 Rysunek 2 ...
Страница 978: ...66 NÁVOD K POUŽITÍ CS Obrázek 1 Obrázek 2 ...
Страница 1029: ...52 HASZNÁLATI UTASÍTÁS HU 9 3 IZODÓZIS GÖRBÉK 2D VIZSGÁLATOKHOZ 9 4 IZODÓZIS GÖRBÉK CBCT VIZSGÁLATOKHOZ Vain 3D laitteelle ...
Страница 1043: ...66 HASZNÁLATI UTASÍTÁS HU Kuva 1 Kuva 2 ...
Страница 1108: ...66 INSTRUCŢIUNI DE UTILIZARE RO Figura 1 Figura 2 ...
Страница 1173: ...66 КЕРІВНИЦТВО З ЕКСПЛУАТАЦІЇ UK Малюнок 1 Малюнок 2 ...
Страница 1174: ......