Cleaning
▍
Cleaning the white calibration plate
If the white calibration plate becomes dirty, it will affect measurement accuracy. If the white calibration plate becomes dirty, it should be
cleaned as described below.
1
Make sure that the MYIRO-9 is switched off.
2
Disconnect all cables (USB cable, LAN cable, AC adapter cable).
3
Open the cover.
4
Loosen the two screws as shown completely so that they are
free of the base.
5
Close the cover.
6
Grasp the ends of the MYIRO-9 and tilt the MYIRO-9 body
backward as shown. The base should stay flat.
7
Wipe the surface of the white calibration plate with a clean, dry,
soft, lint-free cloth or lens-cleaning tissue.
• If the white calibration plate is stained, lens-cleaning fluid may be used to
try to remove the stain.
• Never use benzene or other solvents to clean the white calibration plate.
Doing so may cause discoloration or damage to the plate.
• If a stain cannot be removed, contact your retailer.
8
Grasp the ends of the MYIRO-9 and tilt the MYIRO-9 body onto
the base again.
9
Open the cover
10
Tighten the two screws as shown until snug.
• If the screws are loose when measurements are taken, measurement
accuracy may suffer.
• Do not overtighten.
11
Close the cover and reconnect cables.