19
Calibration Method
Operating Procedure
1
Attach the calibration cap to the limiting aperture.
Align, and turn clockwise
Limiting aperture
2
Wait for the status indicator to light up yellow, and then press the
measurement button. Calibration will be performed.
Memo
The calibration result will be indicated by the color and illumination patterns
(lit/flashing) of the status indicator following calibration. (See page 14)
CAUTION
If the status indicator lights up blue, calibration will not be possible
even if the measurement button is pressed. Perform calibration from
the software being used.
Memo
The calibration cap can be stored on the instrument during
measurement.
Attached for storage
Unattached (not stored)
Calibration cap