Aqua-Check Moisture Tester
MA-26X Manual
Страница 1: ...21 Washington Avenue Scarborough ME 888 293 2121 sales myerstest com www myerstest com Aqua Check Moisture Tester MA 26X Manual ...
Страница 2: ...e of the equipment could cause pressures to exceed the safe level for the apparatus Long sleeves gloves and eye protections should be worn at all times when performing the test Unpacking Set Up NOTE Each Aqua Check is carefully inspected prior to packaging and shipment Immediately unpack and inspect the apparatus for signs of damage once received If any damage has occurred report it immediately to...
Страница 3: ...zed With Pulverizing Balls Hold the Aqua Check horizontal and use an orbital motion to spin the balls around inside the chamber Agitate for 20 seconds and rest for 20 seconds repeat this process three times Periodically check the position of the gauge indicator and continue the cycle until the indicator has stabilized Operating Instructions Continued STEP 8 At eye level hold the vessel horizontall...
Страница 4: ...act between reagent and specimen Insure the vessel is held horizontal when mating with cap Insufficient sample preparation Break cohesive material into finer sizes and use pulverizing balls Inadequate pressure seal Check and clean chamber cap gasket Inspect the vessel cap and stirrup for damage or cracks Defective gauge Gauge needle should rest at zero and move smoothly repeat if defective Low tem...