6.1.7.
Access rights
Change password
Password level 1
Password level 1
covers all parameter settings of the MTP200 and is only enabled for authorised personnel
(e.g. maintenance staff, service technician) to access all configurable settings.
The password may have a maximum of 20 alphanumeric characters and is entered and saved in the
corresponding fields as in the screen.
6.1.8.
Access rights
Change password
Password level 2
Password level 2
secures parameter settings for calibration and the function
Overwrite calibration values.
It
is reserved exclusively for technicians with special knowledge of the device and its functions.
Password level 2
is secured by a password (
5180
) assigned by the manufacturer.
A newly assigned password may have a maximum of 20 alphanumeric characters and is entered and saved
according to the indicated fields as in the screen
6.1.9.
Calibration
Calibrate input
PT-100
A calibration is only necessary for the analogue in- and output signal. Before calibration the parameter set
must be read from MTP200 into WINSMART-program. In addition, you have to select the resistance
measurement input for the Pt 100 calibration (see section 10.1 Measurement inputs).
No entry with the
first assignment of a
password!
Enter
5180
for the
first change of the
pre-occupied
password!