10.5.
Analog output
Fig. 24
Parameterization of the analog output:
Range begin
-
0 %
- value analog output signal
Range end
-
100 %
- value analog output signal
Lower-limit
-
minimal output signal
Upper-limit
-
maximal output signal
Value
-
output signal in case of no connection to flow sensor
Filter
-
Filter 1
st
order, adjustable between 0.1 … 99.9 s
Calibration 4mA/2V
-
After pushing the button
adjust
calibration
the analog value-1 can be adjusted to 4.000 mA or
2.000 V and
Calibration 20mA/10V
-
the analog value-2 to 20.000 mA or 10.000 V.
Changed parameters will be lost by leaving the screen mask. They will be transferred and stored in the
MFI3000 by the instruction
Extras/Program parameters
.
Analog output with constant current:
DIP-switch 3 OFF!
Analog output with voltage signal:
DIP-switch 3 ON!