6.1. DIP-Switch Configuration
The output signal can be configured via a DIP switch behind the front panel. After carefully removing
the front panel, the following switch positions can be used to make settings for the RS485 interface
and analog output:
Fig. 7
Switch 1/2:
Switch to ON for RS485 interface termination.
Switch 3:
Defines the analog output mode
4-20 mA DC current (Schwitch 3 OFF) or
0-10 V DC voltage (Switch 3 ON)
Switch 4: not connected
6.2.
Transmitter
MFI3000
Block Diagram
Fig. 8