9. Troubleshooting
Note:
Before continuing, inspect all accessories for any defects.
Table 5 - Troubleshooting Guide
Symptom
Possible Cause
Corrective Action
The console is not
working.
No electricity
Inspect the electrical wall outlet.
DC power adapter
Verify that the DC power adapter cable is
connected to the DC adapter socket on the
console, and the DC power adapter is
connected to the 100-240 Volt wall outlet.
DC power adapter
cable
Examine the cable for any defects.
The console starts
working and stops
immediately.
The air cannot move
through the garment
hose.
Examine hoses for kinks, twists and folds.
One garment inflates but
the second one does not.
The second garment
does not receive air.
Examine its hoses for kinks, twists and folds.
The console stops
working, the
Status
Indicator
turns On in
orange, and the alarm
beeps.
Hose is not connected
properly to garment or
console, or prong plug
is not inserted into
unused air outlet.
Examine and fasten all air connections.
When treating only one limb, always plug
the unused air outlet with the prong plug
provided with the console. If all air
connections are OK and problem persists,
contact Mueller Sports Medicine.
The
Status Indicator
remains ON in orange
Malfunction
Contact Mueller Sports Medicine.
The console works at a
very low pressure,
regardless of the pressure
set by the user.
Defective garment
Replace garment and check again.
Internal malfunction
Contact Mueller Sports Medicine.
An irregular noise.
Console transferring
vibrations to a surface
Make sure the console is standing evenly on
all four of its bumpers.
Internal malfunction
Contact Mueller Sports Medicine.