MTT MS3700DV Скачать руководство пользователя страница 1

 

User's Manual 

MS3700DV 

Slim Plug-In Voltage Divider 

Document No: MQDDK-121130-3                                                                    Rev. 1.30 

 

 

 

AD Ichigaya Building 4F 
2-5 Ichigaya-Honmuracho 
Shinjuku-ku, Tokyo, 162-0845 
Phone: +81 (0) 3-6280-8635 

 

Thank you very much for employing MTT products. 
Upon receipt of the product(s), please check the affixed label to 
make sure that the specifications shown therein conform to your 
requirements. If you find any discrepancies and/or any damage on 
the product(s), please contact MTT or its local representatives. 
You can be assured that all the MTT products are manufactured 
according to the strict quality control standards.

 

 

1. Introduction

 

For proper operation of the instrument(s), we recommend that you 
carefully read this manual prior to operation. The manual should be 
retained for future reference. Please also refer to the relevant 
product specification sheet.

 

 

2. For Safe Operation

 

Be sure to observe the following safety clauses in your operation of 
the instrument(s). MTT cannot undertake any responsibility and 
guarantee for any damage and/or loss caused by improper 
operation contrary to or neglecting these clauses.

 

 

 

 

 

3. Precautions

 

Be sure to observe the following precautions to ensure proper 

operation of the instrument(s). If these precautions are disregarded, 

the performance of the instrument(s) cannot be guaranteed.

 

For Handling:

 

  This is a precision instrument. Do not drop or throw the 

instrument. 

  This product contains electronic parts. Do not splash water on 

the product and do not dip it in water. For installation, select a 

place where no condensation occurs. 

  Avoid storing or installing the instrument in locations subject to 

direct sunlight, high temperature, dust, high humidity or vibration. 

For Installation:

 

  The instrument is intended for indoor installation. 

  Refer to section 8 “Mounting / Demounting” to install the unit on 

a DIN rail or wall. 

  Operating environmental conditions: 

- Temperature range: -5 to 55°C 

- Humidity: 5 to 90% RH 

- Altitude up to 2000 meters 

  Do not block the vents in the instrument. 

  The instrument should be installed as shown in the sketch below, 

namely, in such a position that the label on the front panel is 

readable in the right direction. 

  For effective heat dissipation, allow a space of at least 100 mm 

above and below the unit. 

The unit should not be installed side by side having no space 

with any other units. There should be a clearance of at least 10 

mm on each side. 

When units are used in multiple stacks, they should be spaced 

at least 130 mm apart from each other. If the required minimum 

space is not available, take appropriate heat protection 

measures by, e.g., placing a partition between the units. Provide 

a ventilation hole or install a cooling fan to improve airflow. 

 

 

Disregard of this clause may cause fatal 

harm or serious bodily injury due to fire or 

electric shock.

 

 

Disregard of this clause may cause bodily 

harm or damages to nearby objects due to 

electric shock or other accidental 

happenings.

 

 

 

  All electrical connections must be made to the terminal block of 

the socket. All connections and disconnections must be made 
with no signal input applied. Otherwise, electric shock may 
result. 

  Do not disassemble or modify the instrument, and do not replace 

the power fuse. Otherwise, fire or electric shock could result. 

  Do not allow any foreign matter (metal chips, water, liquid, etc.) 

to get into the instrument. Should any foreign matter enter the 
instrument, immediately unplug the power cable, and contact 
MTT or its distributor. 

  When the instrument is used for applications that require higher 

reliability and safety, such as transportation, communication, 
power generation control and medical equipment, special 
consideration should be taken in safety design to prevent such 
equipment from malfunctioning as a whole system. 

 Do not use the instrument in an explosive atmosphere, 

containing flammable vapors, gases or dusts. Otherwise, an 
explosion may result. 

  Do not place any combustible materials in the vicinity of the 

instrument. 

  Do not insert or remove the unit with power applied. Otherwise, 

electric shock may result. 

 

 

  Use the instrument within the operating conditions described in 

the published product specification. Failure to do so may cause 

fire or damage to the instrument. 

  Avoid operating the instrument in locations where extreme 

temperature changes can cause condensation. Otherwise, it 

may be damaged and malfunction. 

  Avoid operating or storing the instrument in locations where 

corrosive gases are present or chemical solvents may splash. 

  For the sake of safety, installation and wiring must be performed 

by qualified personnel with expertise in electronics, electricity or 

instrumentation engineering. 

CAUTION 

 

WARNING 

 

WARNING 

 

CAUTION 

 