background image

All specifications are subject to change. Contact MTS for specifications and 
engineering drawings that are critical to your application. Drawings contained 

in this document are for reference only. Go to http://www.mtssensors.com for 
the latest support documentation and related media.

M-Series Model MR Transmitter Electronic Module Removal and installation instructions

Level Plus

®

Magnetostrictive Liquid-Level Sensors

with 

Temposonics

® 

Technology

 M-Series Model MR

Transmitter with Analog Output

Electronics Replacement Instructions

SENS

O

RS

®

Document Part Number

550731 Revision E

The Level Plus

®

 Model MR level transmitter is modular in design. The 

electronics module can be replaced in the field without on site MTS 
applications engineering support.

Before you begin

•  Ensure that all power is disconnected 

•  Follow all lockout procedure(s) are followed prior to opening 

    the transmitters instrumental housing.

To replace the M-Series electronics module, complete the following 
steps:

1.  Remove any dirt, debris, or liquid from the top of the instru

 ment 

enclosure.

2.  Remove the instrument housing cover.

3.   Remove the existing transmitter electronics module, or puck, 

 

by grasping the perimeter of the puck and pulling it firmly 

 

upwards away from the base.

4.  Install the new puck by aligning the connectors at the bottom 

 

of the puck with the appropriate connectors in the base of the 

 

instrument enclosure. Using even pressure on the face of the 

 

puck, press down firmly. Check that the puck is replaced 

 

correctly, by pressing down around on the perimeter of the 

 

puck face. The puck should feel secure and not “rock”.

5.  Refer to the M-Series Operation and Installation Manual, part 

 

number 550720, (available at www.mtssensors.com)  for 

 

calibration and setup instructions.

6.  Replace the instrument housing cover.

Loop 1 Test

    +     —

Loop 2 Test

    +     —

HART

T

O

R

E

M

O

V

E

EL

EC

TR

ON

ICS

M

OD

ULE,

PULLU

NIT IN UPWARD D

IRE

CT

IO

N

- D

O

N

O

T

TW

IS

T

O

R

TU

R

N

Loop 1 test port 

Loop 2 test port

HART test port

Up Key

Enter Key

Down Key

Custom LCD Display

Level Plus

®

M-Series Transmitte

r

Model MR Rigid Transmitter
Single-Cavity Housing