Running the Example HCF Test
Landmark™ Test System Operation
91
2. Adjust force limits so that they are just outside expected values for the HCF test.
A. From the Station Manager main window, select
Display > Station Setup
, and click
the
Channel Input Signals
button.
B. Select the
Force
channel.
C. Select the
Limits
tab.
D. Use the following parameters:
l
Landmark 1 Force (Upper Limit, Upper Action)
— 15000 N, Interlock
l
Landmark 1 Force (Lower Limit, Lower Action
— (-1000) N, Interlock
Show Station Manager Scope and Meters for the Test
1. It is common practice to use the Scope and Meters from the Station Manager application to
show what is happening on the physical equipment during the test. Ensure that the Scope and
Meters you set up earlier are open and running.
Note:
Your meter readings will reflect your station and will be different from those
shown.