MTS 793 Controller Hardware
210
Cable specification
•
15 contact type D male EMI connector.
•
Backshell–EMI metallized plastic.
•
Cable–24 AWG 4 connector with braided shield, with the braid
connected to a metallized plastic backshell at the chassis and to
ground at the emergency stop station.
Jumper plug required
If connector
J29
is not used, you must install a jumper plug to
maintain the integrity of the interlocks. Use jumper plug PN 100-007-
947B or jumper pins: 3-4, 5-7, 8-13.
J43 Interlock
Connector
J43 Interlock
contains one optically isolated interlock
input and one relay contact interlock output.
Note
The intlk out relay (Pin 6) is normally energized when no interlocks
are active. So even though Pin 6 is a NO (normally open) relay
contact, it is actually normally closed when no interlocks are active.
Similarly, the intlk out relay (Pin 8) is normally closed when no
interlocks are active, so even though Pin 8 is a NC (normally closed)
relay, it is actually normally open when no interlocks are active.
5
Local E stop
Local E stop loop (connected to pin 7)
6
N/C
No connection
7
Local E stop
Local E stop loop (connected to pin 5)
8
HPU E stop
E stop loop from HPU (connect to pin 13)
9
N/C
No connection
10
N/C
No connection
11
N/C
No connection
12
xhead power
+24 V crosshead unlock power output (from HSM high relay
contact)
13
HPU E stop
E stop loop from HPU (connect to pin 8)
14
xhead power return
24 V common for crosshead unlock power output
15
N/C
J29 Connector Pin Assignments (part 2 of 2)
P
IN
N
AME
F
UNCTION