Levelling the Main Frame
MTS Exceed™ 22 Series Product Information
Installation
27
Levelling the Main Frame
After installation is ready, you need to level the tester immediately. This can
prevent the swinging of its bottom and therefore provides a level test surface to
enable more accurate test results.
Tools
The following tools are required:
•
Level meter
•
Wrench
Process
Level the base of the main frame.
1. Put the level meter horizontally on the base plane of the main frame.
2. Loosen the locknut of each foot.
3. Use the monkey wrench to adjust the height of each shim in turn and check
readings of the level meter until the levelness is 0.2/1000.
4. Rotate the level meter by 90° to check whether the two sides and front and
back are level.
5. After the equipment is level, tighten the locknut of each shim.
Anchor bolts
Adjusting screw
Shim
Locknut
Main frame
Содержание Exceed 22 Series
Страница 1: ...Manual Title Additional Information be certain Product Information 100 302 290 A MTS Exceed 22 Series...
Страница 24: ...MTS Exceed 22 Series Product Information 24 Dimensions Introduction E22 shield 787mm 2238mm 2145mm...
Страница 54: ...MTS Exceed 22 Series Product Information 54 Shortcut Operations Operation...
Страница 58: ...MTS Exceed 22 Series Product Information 58 Troubleshooting...
Страница 60: ...MTS Exceed 22 Series Product Information 60 Equipment Disassembly...
Страница 67: ......
Страница 68: ...m MTS Systems Corporation http www mts com en Global index asp ISO 9001 Certified QMS...