5.7 Technical data
Level Output
Measured Variable
Product level and interface level
Output Signal /Protocol
Modbus RTU, DDA, Analog (4…20 mA), HART
®
Order Length
1575…22000 mm (62…866 in.) (order length equals the measurement range plus the inactive zone /
contact factory for longer lengths)
Inherent Accuracy
±1 mm (0.039 in.)
Repeatability
0.001% F.S. or 0.381 mm (0.015 in.) whichever is greater
(any direction)
Temperature Output
Measured Variable
Average and multipoint temperatures (Modbus, DDA)
Single point temperature (Analog, HART
®
)
Temperature Accuracy (Modbus, DDA)
±0.2 °C (0.4 °F) range −40…−20 °C (−40…−4 °F),
±0.1 °C (0.2 °F) range −20…+70 °C (−4…+158 °F),
±0.15 °C (0.3 °F) range +70…+100 °C (+158…+212 °F),
±0.5 °C (0.9 °F) range +100…+105 °C (+ 212 …221 °F)
Temperature Accuracy (Analog, HART
®
)
±0.28 °C (0.5 °F) range −40…+105 °C (−40…+221 °F)
Electronics
Input Voltage
10.5…28 VDC
Fail Safe
High, Full scale (Modbus, DDA)
Low, 3.5 mA default or High, 22.8 mA (Analog, HART
®
)
Reverse Polarity Protection
Series diode
EMC
EN 61326-1, EN 61326-2-3, EN 61326-3-2, EN 61000-6-2, EN 61000-6-3, EN 61000-4-2, EN 61000-
4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, EN 61000-4-11
Environmental
Enclosure Rating
NEMA Type 4X, IP65
Humidity
0…100% relative humidity, non-condensing
Operating Temperatures
Electronics: −40…+71 °C (−40…+160 ºF)
Sensing element: −40…+125 °C (−40…+257 °F) (contact factory for specific temperature ranges)
Temperature element: −40…+105 °C (−40…+221 °F)
Vessel Pressure
Flexible Hose: 30 bar (435 psi)
Materials
Wetted parts: 316L stainless steel (contact factory for alternative materials)
Non-wetted parts: 316L stainless steel, Epoxy coated aluminum
Field Installation
Housing Dimensions
Single cavity: 145 mm (5.7 in.) W × by 127 mm (5 in.) D × 109 mm (4.3 in.) H
Dual cavity: 117 mm (4.6 in.) W × by 127 mm (5 in.) D × 206 mm (8.1 in.) H
Stainless steel single cavity: 178 mm (7.1 in.) W × by 135 mm (5.3 in.) D × 153 mm (6 in.) H
NEMA Type 4X: 87 mm (3.4 in.) W × by 124 mm (4.9 in.) D × 132 mm (5.2 in.) H
Mounting
Flexible hose
1 in. Adjustable MNPT or BSPP fitting, Flange mount
Wiring
Connections
4 wire shielded cable or twisted pair,
4570 mm (180 in.) integral cable with pigtail
Daniel Woodhead 6 pin male connector
Electrical connections
Single and dual cavity
¾ in. FNPT conduit opening, M20 for ATEX/IECEx version
NEMA Type 4X
½ in. FNPT conduit opening
Display
Measured variables
Product level, interface level and temperature