55
or/and autoclave and then lubricate socket with the petroleum jelly.
In case of surface damage, crevice or other change, as well as
corrosion, the part (rotor, bucket, etc.) should be immediately
replaced.
Clamping rotor, containers and reducer inserts must be cleaned regularly to
prevent corrosion.
Cleaning of the accessories should be carried out outside of the centrifuge at
least
once every week
if not
after each use. Use a neutral agent of pH value
6÷8 for cleaning the accessories. It is forbidden to use alkaline agent of
pH >
8
. Parts should then be dried using soft fabric or in the chamber drier at ca.
50
°
C.
Angle rotor should be placed on a fabric with holes facing down, for effective
drying.
Do not use bleach on plastic parts of the rotor.
If maintained this way, the useful service life of the device is substantially
increased and susceptibility to corrosion is diminished. Accurate
maintenance increases the service life and protects against premature rotor
failures.
Do not use bleach on plastic parts of the rotor.
According to laboratory standards, minimise the immersion time in each
solution.
Parts made of aluminium are especially prone to corrosion.
Corrosion and damages resulting from insufficient maintenance cannot be
subject of claims lodged against the manufacturer.
The unused rotor should have the lid removed.
HS
accessories maintenance.
Check the general condition of seals.
Make sure that rubber O-rings are lightly coated with silicone grease. Use
high vacuum grease, e.g. type „C” by LUBRINA.
The rotor pins should be always lubricated with
petroleum jelly
.