Settings Parameters
GB
Model 9010/9020 SIL
5.8
Access Code 53 - Default Settings
Default Factory Configuration
Step
Function
P1
Load Factory Default Settings
"def" symbol is displayed if menu is activated. Press ENTER to load factory default values.
P2
Load User Default Settings
"defC" symbol is displayed if menu is activated. Press ENTER to load user default values.
Access Code
Description
Default Value
Overrange Function (EEEE)
A (Enabled)
-
Positive Dead Band Zero Value
+2 %
-
Negative Dead Band Zero Value
-2 %
-
Horn Output Status
dE (Normally not energized),
AC1 P2
Span Calibration Gas Concentration
50 % FS
AC1 P6
Time Interval Between Calibrations Notifica-
tion
0 (Disabled)
AC1 P8
Detector Sensitivity Reduction Warning
nA (Disabled)
AC2 P1
Time out function
A (Enabled)
AC2 P2
Analog Output signal (during a calibration
procedure)
2 mA
AC2 P3
Alarm Inhibited
A (Alarms active)
AC2 P4
Undervoltage of External 24 VDC
nA (Disabled)
AC2 P5
Undervoltage of External 115/230 VAC
A (Enabled)
AC4 P1
Detector Supply Current Setting
310 mA
AC4 P2
Minimum 4-20 mA Transmitter Supply
Current
25 mA
AC4 P3
Maximum Negative Drift
-3%
AC4 P4
Measuring Units
LEL
AC4 P5
Offset Value
0
AC4 P6
Full Scale Value
100
AC4 P7/ P11/ P15 Alarm Direction
C / W / A - U (Rising alarm)
AC4 P7/ P11/ P15 Alarm Thresholds Values
C=15, W=15, A=30
AC4 P8/ P12/ P16 Digital Outputs Status
C / W / A - E (Energized)
AC4 P9/ P13/ P17 Alarm Reset mode
C=1A, W=1A, A=1M
AC4 P10/ P14/ P18 Alarm Mode
norM (no STEL or TWA)