Safekeeping and Storage
G1 Powered Air-Purifying Respirator (PAPR)
US
14
Safekeeping and Storage
NOTICE
Do NOT put the device in storage with a battery installed. If the battery is installed during storage,
corrosion at the battery terminals can occur and cause permanent damage to the battery and blower
assembly.
Make sure that the device operates correctly before putting it in storage.
(1) Install a battery in the assembled device. Refer to Section 6.2 "Installing the Battery" for instruc-
tions.
(2) Do an airflow test with and without cartridges/canisters. Refer to Section 6.6 "Airflow Testing
without Air Filtration Cartridges/Canisters" or Section 6.8 "Airflow Testing with Air Filtration
Cartridges/Canisters" for instructions.
(3) Do the procedures in Section 4 "Visual Examinations". Replace any component that is damaged.
(4) Remove the battery.
14.1 G1 Facepiece
WARNING!
To prevent damage to the facepiece, do NOT keep loose objects in the facepiece container.
Failure to obey this warning can result in serious injury or death.
Keep the facepiece in a pouch or container.
Do not distort the facepiece during storage.
MSA rubber products are protected by an anti-aging agent that can become visible as a light coating.
This coating is harmless and can be removed during cleaning.
To extend the life of rubber components, keep them in a cool, dry place that is protected from ultravi-
olet radiation, according to ISO 2230, Rubber Products – Guidelines for Storage.
14.2 C420 Respirator
Install the green protective plugs in the inlet and outlet ports of the blower assembly.
Install the green protective plugs in the inlet and outlet ports of the breathing tube.
Keep the respirator in cool, dry, clean ambient air away from sunlight, heat, and moisture.
Keep this operating manual with the respirator.