Maintenance and Recertification
G1 Powered Air-Purifying Respirator (PAPR)
US
13
Maintenance and Recertification
MSA recommends that the G1 PAPR be recertified by MSA every 3 years. Recertification includes
inspection of the blower assembly, replacement of the main seal and filter gasket, and verification of
system performance.
Make sure that this device is examined and serviced regularly by an MSA trained repair technician.
MSA is liable only for maintenance and repairs performed by MSA.
Use only genuine MSA replacement parts.
Changes to devices or components are not permitted and will result in unapproved configurations.
Keep detailed inspection and service records.