User Manual
VARIO
luxx
62 / 129
MRU GmbH, D-74172 Neckarsulm
A window appears
►
Set the desired CO threshold value.
►
Touch “OK”.
The setting is saved.
The stored ppm value determines the CO shutdown threshold. If the
measured value reaches this threshold, the second fresh air purge pump
starts and the CO sensor is separated from the gas path by a valve. If the
value falls below 20 % of the threshold value during flushing, the CO sen-
sor is again supplied with sample gas.
Starting the measurement
►
Touch the menu Measure.
The measurement window appears.
The measurement starts with the set parameters.
Storing measured values
You can assign the measured values to the corresponding site and save
them.
►
Touch the menu Measure.
The measurement window appears
►
Touch the “Context menu” (19).
A selection list appears.
►
Touch „Store “
.
A selection list appears.