MRU GmbH
SWG 100 User Manual
Page
95
of
141
10.4.
Troubleshooting
Error
Possible causes
Solutions
Display does not work.
- System is not power-up.
- connect main power
supply.
- power up automatic
fuse.
- contact customer
support.
Analyzer powers down by
itself.
- The fan is defective.
- Filter mat from the fan is
clogged.
- The protection switch has
been triggered.
- exchange the fan.
- exchange the filter mat.
- Internal short circuit.
Contact an electrician to
determine cause of
defect.
Gas cooler unit does not
work.
- The device is not ready,
after a new start of the
analyzer.
- The ambient temperature is
too hot.
- The gas cooler unit is
defective.
- wait until the gas
cooler-unit is ready.
- be sure, that ambient
temperature is within
operating range.
-Contact customer
support.
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...