INE-
DZB-712
Portable
Multi-parameter
Analyzer
9
d). Press F1 three times and F2 once,
when top of ‘F3’ number is ‘6’, press F3 input digit ‘6’
e). Press ENTER/MENU to finish data input of ’12.6’
Note:
For all the operating instructions below, the operation of function key in correspondence
to F1, F2 and F3 will be operated according to indication and not be operated according
to actual pressing key. For example, ‘presently press F3 to start measurement’ will be
amended to ‘presently press START key to start measurement’.
In correspondence to ‘ENTER/MENU’ key, it is operated according to actual function. If
it is confirmation function, you can press ‘ENTER’ and if it is the menu function, you can
press ‘MENU’.
3.4. Accessories and selective accessories
We’ll introduce several important, necessary accessories below. They are indispensable apparatus
to guarantee some necessary functions.
Meter has IP65 protection class, so all electrode system is made
up of electrode connection water-proof system and actual
measurement electrode(6). Electrode connection water-proof
system is made up of union cap (1), top ring (2), cable obdurate
ring (3), measurement gland cover (4) and aerial plug (5).
4. Operation
4.1. Installation
a. Connection of Electrode System
In order to ensure IP65 protection class, meter has relevant water-proof parts, the installation of all
measurement electrode is as follows:
Insert correspondent measurement electrode in relevant measurement electrode socket on rare
panel of meter, then install in sequence the measurement gland cover, cable obdurate ring, top
ring and union cap, etc…