background image

MUNI User’s Guide 

12 

 

 
 

5.4 Calibration and Bump Testing 

Use this menu to perform zero or span calibration for one or more sensors, bump test the sensors 
and alarms for function, and change the span gas concentration.   

The MUNI should be calibrated on the first day of use and at regular intervals not to exceed 180 
days, depending on use and exposure to particulates, contaminants and sensor poisons. A daily 
bump test should be performed to ensure a functional response of all sensors and alarms.   

 

BUMP TEST

 is defined as a brief exposure to sensor gases, typically 10-20 seconds, just 

long enough to indicate that the sensors are responsive and the alarms are functional, without 
concern for a quantitative measurement. 

 

CALIBRATION

 is defined as exposing the sensor(s) to a known concentration standard gas 

for the full calibration time (typically 30-60 seconds, but longer for HCN) and setting the 
reading of the sensor(s) equal to the concentration of the calibration gas. 

 

Calibration intervals and bump test procedures may vary due to sensor type, ambient conditions, 
local regulations and/or the user’s company policies.* Automatic reminders for calibration and 
bump tests can be set up in the CAL? Menu or using the mPower Suite software (see Section 
6.1). When a calibration or bump is due, a small icon is displayed next to the sensor name: 

 

for Cal and    for Bump as shown below:     
 

   

 
Calibration is also required if:   

 The sensor has failed in a previous bump or calibration   

 The sensor module has been replaced with one whose calibration is overdue.

 

 A new sensor is installed. 

 
For more on calibration frequency see Tech/App Note 3 “How Often to Calibrate Gas Detectors”.*     

 

* The calibration frequency must be defined by the user’s company policy because each application is different and 
may cause a sensor’s sensitivity loss for various reasons out of mPower’s control, such as liquids, dirt or corrosion 
preventing gas from reaching a sensor, or exposure to chemicals that poison a sensor’s function.    Exotic gas sensors 
tend to need more frequent calibration than common O

2

, LEL, CO and H

2

S sensors.    In general we recommend a 

bump test before each day’s use to test sensor response and alarm function.    A calibration check can be performed 
by applying a known concentration gas to see if the sensors still respond within typical limits.    Cal check intervals 
can be increased as the user gains experience in the application.    If a bump or cal check fails, the instrument should 
be given a full calibration.    We recommend no more than one month between full calibrations, but this can be 
extended for up to 6 months if company policy allows. 
 
 

Содержание MUNI MP420

Страница 1: ...MUNI Multi Gas Detectors MP420 User s Guide Rev 1 02 October 2020...

Страница 2: ...Exiting Configuration Mode 11 5 3 Navigating Configuration Mode to Edit Parameters 11 5 4 Calibration and Bump Testing 12 5 5 Sensor Enable Disable 16 5 6 Sensor Alarm Settings 16 5 7 Monitor Set up 1...

Страница 3: ...RATIC READING MAY INDICATE A GAS CONCENTRATION BEYOND UPPER SCALE LIMIT WHICH MAY BE HAZARDOUS ATTENTION DES LECTURES SUP RIEURES A L CHELLE PEUVENT INDIQUER DES CONCENTRATIONS EXPLOSIVES TOUTE LECTUR...

Страница 4: ...ons for Safe Use The MUNI multi gas detector must be calibrated if it does not pass a bump test when a new sensor has been installed or at least once every 180 days depending on use and sensor exposur...

Страница 5: ...ining the user 1 1 Main Features Various models for basic confined space entry CSE compliance plus SO2 and HCN sensors Diffusion sampling Compact robust lightweight wearable Auto backlit LCD with larg...

Страница 6: ...ommunication with mPower Suite configuration and data transfer software The mPower USB cable P N M011 3003 000 is required for a PC to recognize the instrument and communicate with mPower Suite 2 2 Ba...

Страница 7: ...tal Display LCD six alarm LEDs one buzzer and one vibration alarm 3 1 Displays and Keys The LCD provides visual information that includes real time gas readings sensor types event log status battery s...

Страница 8: ...brated no sensor is overdue for a bump test or calibration according to the intervals configured on the instrument Calibration overdue Bump overdue 3 1 2 Keys and Interface The MUNI has two keys These...

Страница 9: ...NI alarms when the battery voltage is low and in other fault conditions Alarm Types and Priority Alarm Type Red LED Buzzer Vibrator Max Sensor Alarm Sensor response limit 3 Flashes sec 3 Beeps sec 1 V...

Страница 10: ...is either not installed or turned off the display shows only the installed active sensors 4 4 Alarm Testing and Backlight Under normal operation mode and non alarm conditions the audible buzzer alarm...

Страница 11: ...ation Gas and Measurement Gas and its Correction Factor are displayed these can be changed in mPower Suite The last display before returning to Real time readings is Comm Mode Pressing Left OK stops t...

Страница 12: ...entered use the key to move the cursor among four digits and press OK to change the input 5 2 Exiting Configuration Mode To exit scroll through the main Configuration Mode Menu using the Right key unt...

Страница 13: ...g the mPower Suite software see Section 6 1 When a calibration or bump is due a small icon is displayed next to the sensor name for Cal and for Bump as shown below Calibration is also required if The...

Страница 14: ...ally methane but propane and pentane are also commonly used The measurement gas is also selected in mPower Suite Correction Factors CFs are calculated and automatically applied to make the display rea...

Страница 15: ...LPM regulator range 0 3 1 0 LPM Press Left OK and wait for the 30 second count down timer and PASS or FAIL calibration result To abort press Right at any time during the count down Turn off the gas su...

Страница 16: ...ue Move the cursor to the and press Left OK to save Repeat with other sensors as needed 5 4 2 4 Single and Multi Gas Bump Tests A bump test is a quick check if the sensors and alarms are functioning w...

Страница 17: ...c alarm sub menu Scroll Right until the desired sensor is flashing and press Left OK to select the sensor alarm e g HI ALM 20 0 Press Right to highlight the desired digit and Left to change the alarm...

Страница 18: ...ice Connected list Click on the S N to get the configuration file from the instrument Edit the configuration parameters as desired including those under the tabs for each sensor in the upper right Cli...

Страница 19: ...th Backlight Automatic turns on automatically under dim light conditions and Manual requires the user initiation by pressing either the Left OK or Right keys Policy Check o Must Bump Cal locks the ins...

Страница 20: ...ent Gas can be selected from a list of over 50 compounds for the Pellistor LEL sensor If Measurement and Cal Gas are different a correction factor is calculated and applied to make the sensor display...

Страница 21: ...s always on and many files can be generated The event files will appear under the Datalog tab on the top of the screen Below is a sample screen of event log information listing event type readings and...

Страница 22: ...Cal Bump Procedures For best results download the latest version of mPower Suite at https www mpowerinc com software downloads Further instructions will be included when the MuniDock becomes availabl...

Страница 23: ...off the front cover 4 Sensors Carefully lift out each sensor to be inspected or replaced Install the replacement sensor Make sure the electrical contact pins are aligned with the holes in the PC boar...

Страница 24: ...ite that buzzer and or other alarms are not turned off Call authorized service center Readings abnormally high or low or noisy Reasons Dirty or plugged filter or inlet Dirty or old sensor Excessive mo...

Страница 25: ...MUNI User s Guide 24 10 Technical Specifications All specifications and listed sensors are subject to change without notice Please check for updates at www mpowerinc com...

Страница 26: ...MUNI User s Guide 25 Technical Support and mPower Contacts mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 Phone 408 320 1266 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Отзывы: