Страница 1: ...mpbio com sampleprep Cat No 116012500 For Research Use Only Protocol Revision 116012500 15OCT Owners Manual ...
Страница 2: ... to keep the original packaging in case you need to return the product for any reason to your distributor or MP Biomedicals Note MP Biomedicals is not responsible for any harm or damage resulting directly indirectly from the use of the Super FastPrep 2 or related product components materials or technologies Some reagents in kits are irritants Wear protective gloves and eye protection and follow th...
Страница 3: ...please see accompanying literature Operation Precautions Operation of the Super FastPrep 2 with a plastic tube at full speed for more than 10 seconds may cause the grinding matrices within the tube to overheat leading to sample degradation and or tube failure If grinding of a sample for over 10 seconds is desired lower speeds should be used and or periodic stopping and cooling of the tube should b...
Страница 4: ...t 2 mL tubes containing lysis matrix The Super FastPrep 2 consists of a unique reciprocating grinding mechanism mounted on a rotary tool It is designed to operate at 600 to 4 800 cycles per minute as long as charge remains in the battery This allows the researcher to maintain control of the speed and time of sample lysis Overall the Super FastPrep 2 is ideal for cordless portable and convenient pr...
Страница 5: ...5 Charging Station Rotary Tool Lithium Ion Rechargeable Batteries 2 Lysing Matrix Tubes Support Plugs Super FastPrep 2TM Unit Voltage converter 220 240 VAC to 100 120VAC ...
Страница 6: ...ace the battery in the charging station by aligning the raised ribs of the battery pack with the slots in the charging station The light on the charging station should flash green while the battery is charging The battery pack is fully charged when the green light stops flashing and remains on Holding the Super FastPrep 2 During operation hold the Super FastPrep 2 as in the picture below ...
Страница 7: ...bases are compatible with the Super FastPrep 1 however support plug included must be installed to help prevent failure of those tubes aOperating Temperatures Optimum 20oC to 45oC Severe cold conditions can drain the battery and affect the number of samples that can be processed with the Super FastPrep 2 Keeping the battery warm e g in the pocket of a jacket or parka prior to use can extend battery...
Страница 8: ... with a conical bottomed plastic tube use the support plug to help prevent failure of the tube Hold the tube and support plug between thumb and index finger Press them together into the holder of the Super FastPrep 2 It is recommended that for best results only Lysing Matrix Tubes supplied by MP Biomedicals be used in the Super FastPrep 2 If you use a 2 mm or greater diameter metal or ceramic ball...
Страница 9: ...s for details concerning sub sequent isolation of DNA RNA Protein from lysate material For a complete listing of FastDNA FastRNA and FastProtein purification and isolation kits please visit www mpbio com sampleprep Cleaning the Super FastPrep 2 Make sure to disconnect the rotary tool from the Super FastPrep 2 before cleaning Wear protective gloves latex nitrile during all cleaning procedures Clean...
Страница 10: ...tered or misused the goods or have failed to use or store them in accordance with instructions furnished by us Our sole and exclusive liability and your exclusive remedy with respect to goods proved to our satisfaction applying analytical methods reasonably selected by us to be defective or nonconforming shall be the replacement of such goods free of charge upon the return of such goods in accorda...
Страница 11: ...11 ...
Страница 12: ...0 629 67 337 India Tel 91 22 22631257 58 Fax 91 22 22631259 Japan Tel 81 03 3808 2102 Tel 0120 78 8020 Fax 81 03 3808 2401 Latin America Tel 1 440 337 1200 Fax 1 440 337 1180 The Netherlands Tel 0800 0227 416 Fax 0800 0227 489 New Zealand Tel 64 9 912 2460 Fax 64 9 838 4209 Poland Tel 48 22 659 58 95 Fax 48 22 658 45 05 Russia Tel 7 495 661 0008 Fax 33 3 88 67 19 45 Serbia Tel 381 11 242 1972 Fax ...