![Movea MotionController Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/movea/motioncontroller/motioncontroller_user-manual_1841144009.webp)
9
VI.
USAGE
1.
RECHARGING YOUR MOTIONPOD™
The MOTIONCONTROLLER™ must be connected to a computer in compliance
with IEC/EN 60950-1 standard. In order to fulfill safety requirements for
electro-medical systems and equipments, power must be supplied by a
transformer in compliance with EN standard 60601-1. Only the MOTIONPOD™ is
in contact with the patient environment. By no means, the patient should be in
contact with the computer and the MOTIONCONTROLLER™ during use.
First connect the MOTIONCONTROLLER™ to the USB port of your computer using the provided
cable.
Once the MOTIONCONTROLLER™ is connected, the light indicator “pwr” goes on and gives a green
continuous light (see Figure 2). The two other light indicators are blue diodes which light up along
with the first one once connected to the USB port. They switch off immediately then blink twice to
finally go off completely.
Put the MOTIONCONTROLLER™ horizontally on a table, facet where the logo is imprinted upwards.
You can see the cavity where you can clip the MotionPod in. Insert the MOTIONPOD™ by its Black
bottom cover directly in the cavity. The button of the MOTIONPOD™ must be pointing to the Logo.
The MOTIONPOD™ must completely fit in the cavity.
A clockwise rotation of approximately 30° mechanically locks the MOTIONPOD™ in the cavity and
therefore electrically connects it to the MOTIONCONTROLLER™. A “click” sound indicates that the
system is suitably locked and started charging (if charging is needed). The whole unit must look like
the photograph on Figure 5.
The MOTIONPOD™ encloses a battery. Battery performance gradually decreases
over an extended period of time (see the number of possible charging cycles). In
case of extended period of no use, it is advised to check the battery condition by
charging it. If the battery does not charge any more it is necessary to return the
product to the manufacturer for maintenance (battery exchange). This
operation cannot be carried out by the user. Opening of MOTIONPOD™ (or
MOTIONCONTROLLER™) will void the manufacturer warranty.