SCALE CALIBRATION (MODELS WITH A SCALE ONLY)
4 - 3
Step 1 - Electronic Entry into Calibration Mode
Encompasses all scales sold
without
a mechanical calibration switch installed. When the integrated scale
does not have a calibration switch, electronic entry is required.
Firmware Versions Older Than v3.4
To perform an electronic calibration entry:
•
Touch
Scale Zero
and
Volume/Tone
simultaneously for five consecutive seconds (see
for the location of these buttons on the front panel).
Firmware Versions Newer Than v3.4
To perform an electronic calibration entry:
•
Press and hold the
Scale Zero
and
Volume
buttons continuously for five seconds. After five seconds a
short beep sounds. Release the
Scale Zero
and
Volume
buttons.
•
Within two seconds after releasing the buttons, press the
Scale Zero
and
Volume
buttons again and
release. After release the system sounds five long beeps and enters scale calibration.
When Calibration Mode is entered successfully, the scanner emits five long beep sounds. The optional Scale
Display blinks
CAL00
and
CAL _ _
and the 7-segment display scrolls
C00Lb
, or
C00g
. Once the scale
Calibration Mode is successfully entered, continue the Calibration process with
.
Step 2 - Program Legal Parameters
Sealable and legal parameters (
Unit Selection
and
Dampening Filter Setting
) can only be
programmed/changed when the MP6000 is placed into Calibration Mode. There is no requirement to change
these values (they can remain in the default or present state). However, if there is a need to change their
values they must be programmed when the scale is in a Calibration Mode. The
Dampening Filter Setting
can
be changed multiple times during a Calibration Mode without affecting the calibration outcome.
After completing these changes, the calibration process re-starts with
Step 3 - Calibration at NO LOAD on
. If you change the
Unit Selection
value before successfully reaching
Step 5 - Calibration Success or
and the Calibration Mode exits, the scale will be non-operational. The scale will have to be
successfully calibrated before it becomes operational.
Legal Scale Units (Unit Selection) - Kilograms or Pounds
To change the legal parameter
Unit Selection
(when the scale is in a Calibration Mode):
1.
Verify the units of weight measurement by checking the
Scale Display
icon (the icon displays either
lb
or
kg
units), or the 7-segment display which scrolls either
Lb
or
g
units (
g
indicates kg units).
IMPORTANT
If Calibration Mode was not entered successfully, the scale remains in its current state.
Содержание MP6000
Страница 1: ...MP6000 INTEGRATOR GUIDE ...
Страница 2: ......
Страница 3: ...MP6000 INTEGRATOR GUIDE 72E 172632 06 Revision A November 2014 ...
Страница 7: ...v ...
Страница 8: ...vi MP6000 INTEGRATOR GUIDE ...
Страница 14: ...xii MP6000 INTEGRATOR GUIDE ...
Страница 18: ...viii MP6000 INTEGRATOR GUIDE ...
Страница 33: ...PRODUCT OVERVIEW AND FEATURES 1 15 ...
Страница 34: ...1 16 MP6000 INTEGRATOR GUIDE ...
Страница 41: ...HOST INTERFACES AND CABLE PINOUTS 2 7 USB Device Type continued IBM Table top USB ...
Страница 42: ...2 8 MP6000 INTEGRATOR GUIDE USB Device Type continued IBM Hand held USB ...
Страница 45: ...HOST INTERFACES AND CABLE PINOUTS 2 11 USB Device Type continued CDC COM Port Emulation ...
Страница 55: ...HOST INTERFACES AND CABLE PINOUTS 2 21 RS 232 Host Types continued Standard RS 232 ...
Страница 56: ...2 22 MP6000 INTEGRATOR GUIDE RS 232 Host Types continued ICL RS 232 ...
Страница 57: ...HOST INTERFACES AND CABLE PINOUTS 2 23 RS 232 Host Types continued Wincor Nixdorf RS 232 Mode A ...
Страница 58: ...2 24 MP6000 INTEGRATOR GUIDE RS 232 Host Types continued Wincor Nixdorf RS 232 Mode B ...
Страница 59: ...HOST INTERFACES AND CABLE PINOUTS 2 25 RS 232 Host Types continued Olivetti ORS4500 ...
Страница 60: ...2 26 MP6000 INTEGRATOR GUIDE RS 232 Host Types continued Omron ...
Страница 61: ...HOST INTERFACES AND CABLE PINOUTS 2 27 RS 232 Host Types continued OPOS JPOS ...
Страница 62: ...2 28 MP6000 INTEGRATOR GUIDE RS 232 Host Types continued Fujitsu RS 232 ...
Страница 63: ...HOST INTERFACES AND CABLE PINOUTS 2 29 RS 232 Host Types continued CUTE 2 ...
Страница 72: ...2 38 MP6000 INTEGRATOR GUIDE Third Party Scale continued Disable Third Party Scale 0 ...
Страница 74: ...2 40 MP6000 INTEGRATOR GUIDE Third Party Scale LED Pin continued Active High 1 ...
Страница 76: ...2 42 MP6000 INTEGRATOR GUIDE Third Party Scale Zero Pin continued Active High 1 ...
Страница 79: ...HOST INTERFACES AND CABLE PINOUTS 2 45 IBM Port Addresses None Selected ...
Страница 80: ...2 46 MP6000 INTEGRATOR GUIDE IBM Port Addresses continued Hand held Scanner Emulation Port 9B ...
Страница 81: ...HOST INTERFACES AND CABLE PINOUTS 2 47 IBM Port Addresses continued Non IBM Scanner Emulation Port 5B ...
Страница 82: ...2 48 MP6000 INTEGRATOR GUIDE IBM Port Addresses continued Tabletop Scanner Emulation Port 17 ...
Страница 84: ...2 50 MP6000 INTEGRATOR GUIDE IBM Scale Port Addresses continued Port 6A ...
Страница 85: ...HOST INTERFACES AND CABLE PINOUTS 2 51 IBM Scale Port Addresses continued Port 6B ...
Страница 86: ...2 52 MP6000 INTEGRATOR GUIDE IBM Scale Port Addresses continued Port 6E ...
Страница 90: ...2 56 MP6000 INTEGRATOR GUIDE ...
Страница 130: ...3 40 MP6000 INTEGRATOR GUIDE ...
Страница 144: ...4 14 MP6000 INTEGRATOR GUIDE Legal Scale Units continued Pounds 1 ...
Страница 146: ...4 16 MP6000 INTEGRATOR GUIDE Scale Display Configuration continued Disable Scale Display Configuration 0 ...
Страница 148: ...4 18 MP6000 INTEGRATOR GUIDE Parameter 996 continued Low Vibration Sensitivity 1 ...
Страница 149: ...SCALE CALIBRATION MODELS WITH A SCALE ONLY 4 19 Parameter 996 continued Very Low Vibration Sensitivity 2 ...
Страница 150: ...4 20 MP6000 INTEGRATOR GUIDE Parameter 996 continued Ultra Low Vibration Sensitivity 3 ...
Страница 154: ...4 24 MP6000 INTEGRATOR GUIDE ...
Страница 188: ...A 8 MP6000 INTEGRATOR GUIDE ...
Страница 192: ...B 4 MP6000 INTEGRATOR GUIDE ...
Страница 208: ......
Страница 209: ......