background image

 

16 

 

VIII Care and maintenance 

   

Lenses and filters 

 

 

To clean lens surfaces or filters, first remove dust using compressed air. If dust still persists, use a 

soft/clean brush or gauze. 

 

A soft gauze or lens tissue lightly moistened with pure alcohol should only be used to remove grease or 

fingerprints. 

 

Use petroleum benzine to clean immersion oil. 

 

Use petroleum benzine only to remove immersion oil from objective lenses. 

 

Because petroleum benzine and absolute alcohol are both highly flammable, be careful handling around 

open flame. 

 

Do not use same area of gauze or tissue, to wipe more than once.

 

 

Cleaning of painted or plastic components 

 

 

Do not use organic solvents (thinners, alcohol, ether, etc.).    Doing so could result in discoloration or in 

the peeling of paint. 

 

For stubborn dirt, moisten a piece of gauze with diluted detergent and wipe clean.

 

 

When not in use   

 

 

When not in use, cover the instrument with dust cover and store in a place low in humidity where mould 

is not likely to form.   

 

Store the objectives, eyepieces and filters in a container or desiccator with drying agent. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание SWIFTLINE MAE-31R

Страница 1: ...1 Inverted Microscope Instruction Manual Model MAE 31R...

Страница 2: ...ounting the eyepieces 7 V Microscopic procedure Interpupillary distance adjustment 7 1 Diopter adjustment 8 2 Centering the Condenser 8 3 Centering the lamp 9 4 Brightfield microscopy 11 5 Phase contr...

Страница 3: ...Clamp Screw 9 Revolving Nosepiece 3 Annular Diaphragm 10 Coaxial Coarse Fine Focus Knob 4 Condenser Diaphragm Lever 11 Torque Adjusting Ring 5 Condenser Focus Knob 12 Filter Retaining Ring 6 Condenser...

Страница 4: ...r 6 Filter Slider 7 Lamp House 8 Field Diaphragm Lever 9 Condenser Clamp Screw 10 Phase Slider 11 Condenser Clamp Holder Screw 12 Condenser Centering Screws 13 Stage Plate 14 Fluorescence Filter casse...

Страница 5: ...on N A W D mm Plan Achromat 4X 0 10 23 5 Plan Achromat Phase 10X 0 25 7 5 Plan Achromat Phase LWD 20X 0 40 7 0 Condenser Extra Long Working Distance N A 0 3 W D 72mm Electrical Specifications Input IE...

Страница 6: ...y damage the electrical components of the microscope In order to prevent electric shock always turn the power switch on the power supply off before connecting the power cord 1 Installing the lamp In o...

Страница 7: ...into the nosepiece so that the magnification increases with clockwise rotation of the revolving nosepiece Replace the stage plate insert 4 Mounting the eyepieces Remove the dust caps from the eyepiece...

Страница 8: ...ocus using the 10x objective Turn the diopter compensation ring on each eyepiece until the adjustment ring is adjusted to 0 position Position 40x objective into the optical path and bring the specimen...

Страница 9: ...f view Adjust the condenser centering screws so that the center of the field diaphragm image matches the center of the field of view This adjustment is easier to make if the field diaphragm size is re...

Страница 10: ...low the knob to move around with little resistance Holding the lamp socket clamp screw move the lamp filament image to the center of the condenser diaphragm iris The filament will be visible looking d...

Страница 11: ...t is important because it determines the resolution of the image contrast depth of focus and brightness Closing down the aperture diaphragm will lower the resolution and brightness but increase the co...

Страница 12: ...be and insert the phase centering telescope in its place Loosen the locking screw of the centering telescope eyepiece and slide out until both the phase plate image of the objective and the annular di...

Страница 13: ...e filter in the designated retaining ring above the phase annular diaphragm slider Filter selection Filter type Procedure GIF Green interference 546nm For phase contrast and contrast adjustment with b...

Страница 14: ...mera simply screw onto your C Mount camera Remove the cap on the camera port of your microscope by loosening the knurled screw Then attach the C0 5X and lock into place with the knurled screw Then fol...

Страница 15: ...tfield objective being used Phase annular diaphragm not in optical path Phase annular diaphragm and objective phase symbol do not match Slider annular ring image has moved away from the objective phas...

Страница 16: ...enzine and absolute alcohol are both highly flammable be careful handling around open flame Do not use same area of gauze or tissue to wipe more than once 2 Cleaning of painted or plastic components D...

Страница 17: ...ceipt with the RMA number also clearly written on the exterior of the shipping box Microscope s must be shipped back to originating location or sales office A detailed description of the difficulty mu...

Отзывы: