13 

 

 

If the objective phase plate and the annular of the slider do not coincide, use the two hexagonal 

screwdrivers supplied with the microscope to bring the slider annular ring to the center of the   

phase plate, so that the image of the annular diaphragm is concentric with the phase plate image.

 

 
 
 
       

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

If the slider annular ring image is moved from the phase plate image in the objective, a low   

phase contrast image will result. 

 

For phase contrast microscopy at the maximum contrast, use GIF (Green interference filter)   

in the optical path. 

 

Place the filter in the designated retaining ring above the phase annular diaphragm slider.

 

 
Filter selection 

 

Filter type 

Procedure 

GIF (Green interference) 546nm 

For phase contrast and contrast adjustment with black 
and white film 

NCB (Neutral Color Balance) 
Blue 

For general microscopy and color photomicrography 

 

VI  Photo Procedure 

 

 

The optical path selector lever can be used to set the optical path to either the Binocular tube 100:0 or 

Binocular tube/vertical tube 20:80 (observation: photo). 

 

 

 

Before starting, make sure all prior steps in this manual are followed.   

 

In order to attach a camera to the microscope, an additional camera adapter must be purchased. 

 

 

See picture below for reference. Also refer to the parts and accessories section on page 17. 

Содержание SWIFTLINE MAE-31R

Страница 1: ...1 Inverted Microscope Instruction Manual Model MAE 31R...

Страница 2: ...ounting the eyepieces 7 V Microscopic procedure Interpupillary distance adjustment 7 1 Diopter adjustment 8 2 Centering the Condenser 8 3 Centering the lamp 9 4 Brightfield microscopy 11 5 Phase contr...

Страница 3: ...Clamp Screw 9 Revolving Nosepiece 3 Annular Diaphragm 10 Coaxial Coarse Fine Focus Knob 4 Condenser Diaphragm Lever 11 Torque Adjusting Ring 5 Condenser Focus Knob 12 Filter Retaining Ring 6 Condenser...

Страница 4: ...r 6 Filter Slider 7 Lamp House 8 Field Diaphragm Lever 9 Condenser Clamp Screw 10 Phase Slider 11 Condenser Clamp Holder Screw 12 Condenser Centering Screws 13 Stage Plate 14 Fluorescence Filter casse...

Страница 5: ...on N A W D mm Plan Achromat 4X 0 10 23 5 Plan Achromat Phase 10X 0 25 7 5 Plan Achromat Phase LWD 20X 0 40 7 0 Condenser Extra Long Working Distance N A 0 3 W D 72mm Electrical Specifications Input IE...

Страница 6: ...y damage the electrical components of the microscope In order to prevent electric shock always turn the power switch on the power supply off before connecting the power cord 1 Installing the lamp In o...

Страница 7: ...into the nosepiece so that the magnification increases with clockwise rotation of the revolving nosepiece Replace the stage plate insert 4 Mounting the eyepieces Remove the dust caps from the eyepiece...

Страница 8: ...ocus using the 10x objective Turn the diopter compensation ring on each eyepiece until the adjustment ring is adjusted to 0 position Position 40x objective into the optical path and bring the specimen...

Страница 9: ...f view Adjust the condenser centering screws so that the center of the field diaphragm image matches the center of the field of view This adjustment is easier to make if the field diaphragm size is re...

Страница 10: ...low the knob to move around with little resistance Holding the lamp socket clamp screw move the lamp filament image to the center of the condenser diaphragm iris The filament will be visible looking d...

Страница 11: ...t is important because it determines the resolution of the image contrast depth of focus and brightness Closing down the aperture diaphragm will lower the resolution and brightness but increase the co...

Страница 12: ...be and insert the phase centering telescope in its place Loosen the locking screw of the centering telescope eyepiece and slide out until both the phase plate image of the objective and the annular di...

Страница 13: ...e filter in the designated retaining ring above the phase annular diaphragm slider Filter selection Filter type Procedure GIF Green interference 546nm For phase contrast and contrast adjustment with b...

Страница 14: ...mera simply screw onto your C Mount camera Remove the cap on the camera port of your microscope by loosening the knurled screw Then attach the C0 5X and lock into place with the knurled screw Then fol...

Страница 15: ...tfield objective being used Phase annular diaphragm not in optical path Phase annular diaphragm and objective phase symbol do not match Slider annular ring image has moved away from the objective phas...

Страница 16: ...enzine and absolute alcohol are both highly flammable be careful handling around open flame Do not use same area of gauze or tissue to wipe more than once 2 Cleaning of painted or plastic components D...

Страница 17: ...ceipt with the RMA number also clearly written on the exterior of the shipping box Microscope s must be shipped back to originating location or sales office A detailed description of the difficulty mu...