B. Mechanical 

Maintenance 

 

4.  Adjusting the tension of the coarse focus knob.  

 

The collar to adjust the tension of the coarse focus knob (Fig. 1) is situated between 
the coarse focus knob (9) and the arm (12).  
The tension of the coarse focus is pre adjusted by the manufacturer. The optimum 
tension point is that which permits the coarse focus knob to move as smoothly as 
possible, without the head (7) sliding down by itself. 

 

a.  To adjust tension first loosen the screw situated in one of the four orifices of the 

collar with a 2mm hexagonal key.  

 

b.  To tighten the tension of the coarse focus knob, turn the collar in the clockwise. To 

loosen it, turn anti clockwise.  

 

c. Tighten 

hexagonal 

screw. 

 

5.  Adjusting the rack stop. 

 

The 40X objective use a retractable 
security system to avoid damage to the 
slide or to the front of the lens should they 
come into contact. An additional security 
measure is the security rack stop screw 
that regulates the descending movement 
of the head. It is adjusted by the 
manufacturer for the observation of 
standard slides with a cover slip 0.17mm 
thick; but for the observation of other types 
of slides, readjustment may be necessary.   

 
 

Tensión 

adjustment

collar

Set screw

Screw used to 

adjust the height 

of the head

a.  Firstly, loosen the rack stop 

screw. (Fig. 1) 

 
b.  Next, unscrew the adjustment 

screw. (Fig. 1) 

 

c.  Turn the fine focus knob (10) 

until at mid point (not applicable 
for SFC-3A). 

 

Fig. 1 

d.  Focus on the slide only using the coarse focus control (9) with the 4X objective (5) 

firstly, and then the 10X. 

 

e.  Tighten the adjustment screw only until firm, so that the head (3) cannot descend.  

 

f. 

Tighten the rack stop screw. 

 
 
 
 
 
 
 
 
 
 
 

-6- 

Содержание SFC-3 Series

Страница 1: ...Instruction Manual SFC 3 Series Copyright 05 03 Motic Microscopes European Division...

Страница 2: ...epiece holder tube 10 Fine focus knob 9 Coarse focus knob 12 Arm 8 Diaphragm 6 Stage clips 7 Stage 5 Objectives 4 Revolving nosepiece 1 Eyepiece 3 Head 11 Head adjustment screw 13 Mirror 14 Base SFC 3...

Страница 3: ...All components of the microscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the...

Страница 4: ...ncave mirror 13 Proportions the illumination on the sample Operating A Illumination 1 Place the microscope near a light source or a window 2 Place the microscope in the most comfortable position for u...

Страница 5: ...e model SFC 3A adjust using the coarse focus knob 9 3 When changing to the 40X take care to ensure that the objective does not rub against the slide as damage can occur to the front lens of the object...

Страница 6: ...40X objective use a retractable security system to avoid damage to the slide or to the front of the lens should they come into contact An additional security measure is the security rack stop screw th...

Страница 7: ...e needs adjusting Readjust the stopper Moving the microscope Avoid moving the microscope if possible Carry the microscope in both hands with one hand holding the arm 12 and the other supporting the ba...

Отзывы: