(5) Replacement Parts
* Replace the cooling water once a year.
* Replace the deionization filter cartridge once a year.
* We recommend replacing the flash lamp after it has exceeded 10,000,000 pulses; after this, errors may
occur. After 20,000,000 pulses, the lamp is at the end of its working life and must be replaced; otherwise,
various errors will occur with increasing frequency. See page 34 for how to check the total number of
pulses for the flash lamp by using the Menu.
* Order parts through your local dealer or J. MORITA OFFICE.