Table 3
MANUFACTURER'S GUIDE AND DECLARATION
THE LIFTING ARMCHAIR IS INTENDED TO FUNCTION IN THE FOLLOWING ELECTROMAGNETIC
CONDITIONS. THE CLIENT OR USER MUST ENSURE THAT IT IS USED IN THESE CONDITIONS.
IMMUNITY TESTS
TEST LEVEL IEC 60601
CONFORMITY LEVEL
ELECTROMAGNETIC
ENVIRONMENT – GUIDE
CONDUCTED
DISTURBANCES
INDUCED BY RF
FIELDS 61000-4-6
3 V EFF.
FROM 150 KHZ
TO 80 MHZ
3 V EFF.
FROM 150 KHZ
TO 80MHZ
PORTABLE AND MOBILE RF
COMMUNICATIONS DEVICES
SHOULD NOT BE USED NEAR
ANY PART OF THE LIFTING
ARMCHAIR, INCLUDING
THE CABLES, UNLESS THE
RECOMMENDED SEPARATION
DISTANCES ARE OBSERVED,
CALCULATED BY THE
EQUATION APPLICABLE TO THE
TRANSMITTER FREQUENCY.
RECOMMENDED DISTANCES
D = 1.2√P
D = 1.2√P
FROM 80 MHZ TO 800 MHZ
D = 2.3√P
FROM 800 MHZ TO 2.5 GHZ
WHERE "P" INDICATES THE
MAXIMUM RATED OUTPUT
OF THE TRANSMITTER IN
WATT (W) ACCORDING TO THE
TRANSMITTER MANUFACTURER
AND "D" IS THE RECOMMENDED
SEPARATION DISTANCE IN
METRES (M). THE INTENSITY
OF THE FIELD GENERATED
BY FIXED RF TRANSMITTERS,
AS DETERMINED IN AN ON-
SITE ELECTROMAGNETIC
INSPECTION, MIGHT BE BELOW
THE LEVEL OF CONFORMITY
IN EACH FREQUENCY
BAND. INTERFERENCE MAY
BE EXPERIENCED IN THE
PROXIMITY OF EQUIPMENT
MARKED WITH THIS SYMBOL:
RF DISTURBANCES
EN 6100-4-3
3 V/M
FROM 80 MHZ
TO 2.5 GHZ
3 V/M
FROM 80 MHZ
TO 2.5 GHZ
NOTE 1: AT 80 MHZ AND 800 MHZ, APPLY THE SEPARATION DISTANCE FOR THE HIGHER FREQUENCY BAND.
NOTE 2: THESE GUIDELINES MAY NOT APPLY IN ALL SITUATIONS. ELECTROMAGNETIC PROPAGATION
IS AFFECTED BY ABSORPTION AND REFLECTION FROM STRUCTURES, OBJECTS AND PEOPLE.
A
THE INTENSITY OF THE FIELD FOR FIXED TRANSMITTERS, SUCH AS BASE STATIONS FOR RADIO-
TELEPHONES (CELLULARS AND CORDLESS) AND TERRESTRIAL MOBILE RADIOS, AMATEUR RADIO
EQUIPMENT, AM AND FM RADIO TRANSMITTERS AND TV TRANSMITTERS MAY NOT BE KNOWN
PRECISELY IN THEORY. TO ESTABLISH AN ELECTROMAGNETIC ENVIRONMENT CAUSED BY FIXED
RF TRANSMITTERS, ONE MUST RUN AN ON-SITE ELECTROMAGNETIC INSPECTION. IF THE FIELD
INTENSITY MEASURED IN THE PLACE WHERE THE LIFTING ARMCHAIR IS USED EXCEEDS THE
APPLICABLE RF LEVEL OF CONFORMITY INDICATED ABOVE, NORMAL FUNCTIONING OF THE LIFTING
ARMCHAIR MUST BE MONITORED. IF ABNORMALITIES ARE OBSERVED, ADDITIONAL MEASURES MAY
BE NECESSARY, SUCH AS RE-ORIENTATION OR REPOSITIONING OF THE LIFTING ARMCHAIR.
B
THE FIELD INTENSITY OVER THE FREQUENCY RANGE FROM 150 KHZ TO 80 MHZ SHOULD BE LESS THAN 3 V/M.