Page 14

 

 
 

@BAUD

 

 

 

Will display the BAUD rate in kbyte/sec 

@ENET

 

 

 

Will show the Ethernet Address settings. IP Address, Sub net 

mask and Gateway Address 

@OUTPT

 

 

POS or NEG 

Select the polarity of the pulse output. 

@DSPLR

 

 

HALF, 1_SEC, 1.5_S 

This sets the maximum display update rate to ½, 1 or 1½  second. 

@SERNO

 

 

 

View the serial number of the unit. 

@PASSC

 

 

xxxxx 

Set  this  to  00000  for  no  passcode.  Otherwise  the  user  must 
enter the xxxxx passcode to enter the menu.  

If a user tries to enter too many characters, or enters illegal data, the instrument will respond with 

Err

 and will abort 

the process.  If the command is successfully executed, the instrument will again send the 

OK

 message. 

7.0 

SPECIFICATIONS 

Speed Range 

5 RPM to 999,990 RPM  

(speeds below 5 RPM are possible with multiple pulses per revolution) 

Input Frequency Range 

0.083 Hz to 250 KHz 

Totalizer / Counter 

Display Range: 0.001 to 99,999 

100,000 – 999,990 (only top 5 digits are shown -- x10)  

Maximum # of counts of 16,777,216 even with a 0.001 scale 

Input Configuration 
and 
Voltage Range 

Universal inputs:  1 to 9,999 pulses per revolution 

Front panel push button programmable 

TTL input and 1.1 V to 50 Vac signals – Opt: -1 to -50 Vdc 

Accuracy 

±0.001% of reading or ±1 displayed resolution (Standard Gate) 

±0.01% of reading or ±1 displayed resolution (Fast Gate) 

Resolution 

Fixed Range Mode  

 

Auto Range Mode 

1 RPM (5 to 99,999 RPM) 

 

Up to 0.001 RPM 

10 RPM (100,000 to 999,990 RPM) 

Display 

5 digits 0.56 inch high red seven segment LED 

Display Update 

Menu selectable between: 0.5 second above 120 RPM, 1 Second above 60 RPM, 1.5 Second 

above 40 RPM  

Max Measurement Rate 

Up to 100 times per second (Standard Gate) or up to 1000 (Fast Gate)  

Dependent on input frequency. Input must be > measurement rate 

Scale Mode  
Scaling Computation 

Programmable scaling 0.0001 to 99,999 

Front panel push button programmable 

Decimal Point 

User Selectable (0-3 decimal places max) 

Front panel push button programmable 

Memory 

Maximum and minimum recall from front panel push buttons 

Dimensions 

1/8 DIN by 3.94‖ [100 mm] deep 

Panel cut out: 1.74‖ H x 3/58‖ W [44 mm x 91 mm] 

Input Power 

Standard 100 - 240 Vac (50/60 Hz) 8VA 

Or Optional 12 Vdc or 24Vdc isolated 6W 

Recommended Sensors 

Optical – ROS-W   

 

Infrared – IRS-W 

Optical – ROS-HT-W-25 

 

Laser – ROLS-W 

Optical – ROSM-5W 

 

Laser – SLS-115/230 

Proximity – P5-11   

 

Mag/Amp – MT-190W 

Magnetic – M-190W 

 

Gas Engine - GE-200 

Sensor Power Output 

12Vdc @ 60mA standard, or optional 5Vdc @ 60mA or 24Vdc @ 30mA to sensor 

Pulse Repeater Output 

TTL 

0 to 5 V TTL compatible, one pulse out for each pulse in 

Menu selectable polarity 

Serial Interface 

Bi-directional RS232C, USB or Ethernet Interface 

Analog Output 

Voltage 

(AO)

:  0 to 5 Vdc, 5 mA max load, Isolated or Non-isolated

 

OR 
Current Source 

(IO)

:  4 to 20 mA  (500

 max load with internal 12V Compliance Voltage), 

Isolated or Non-isolated. (1-5V with 250

 load resistor) 

15 bit resolution. Front panel push button programmable for full scale and offset RPM ranges. 

Alarm Capability 

Two alarm set points:  High or Low Alarm Limits 
 

Latching or Non-latching 

Front panel push button or remotely programmable 
Hysteresis and low limit lockout are programmable 
Relays can be set as fail safe. 

Alarm Outputs 

Form C relay contacts, rated 1A at 115 Vac or 230 Vac 

Alarm Reset 

Automatic or manual reset. Front panel push button programmable 

Remote reset via AUXiliary input. 

Remote Programming 

Unit can be remotely programmed using optional Serial Interface. 

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...

Отзывы: