Alarms and service personnel
Alarm conditions
Alarm 1: Rotor cover opened
Physiological alarm condition
None
Technical alarm condition
Opening sensor of the cover activated
Alarm limit
None
Alarm delay
< 1 s
Alarm signal delay
< 1 s
Alarm signal
Visible
Alarm signal description in the display
Door opened!
Audio pause
None
Alarm 2: Package temperature too hot
Physiological alarm condition
None
Technical alarm condition
Temperature sensor reports greater than
43.0°C for the package with the
tumescence solution
Alarm limit
43.0°C
Alarm delay
1 s
Alarm signal delay
< 1 s
Alarm signal
Audible/visible
Alarm signal description in the display
Bag too hot!
Audio pause
None
Workplaces of the service personnel
Any handling of the device assumes exact knowledge of this instruction for use and
conformity with it.
This instruction for use does not replace instructions from the user.
Use of the device is authorised only for professional personnel.