0112-0174 C
45
Appendix A: Instrument Specifications
Performance
Base Configuration
Multiple Detectors
Detector
Low-noise photomultiplier tube with simultaneous photon
counting and analog mode
Same as base
configuration
Spectral
range
380-630 nm
Same as base
configuration
Limit of
detection
< 20 attomol ATP per well
Same as base
configuration
Dynamic
range
> 9 orders of magnitude
Same as base
configuration
Crosstalk
96-well < 10
-6
(black plate),
< 3 × 10
-5
(white plate)
96-well < 10
-6
(black
plate),
< 5 × 10
-5
(white plate)
Heater
temperature
range
A 5° C to 45° C (41° F to 113° F)
Same as base
configuration
Plate
Orientation
Landscape and Portrait
Same as base
configuration
Read modes
Endpoint, Dual Read, Kinetic, Fast Kinetic, Dual Wavelength
Same as base
configuration
Compatibility Base Configuration
Multiple Detectors
Microplates
All 96 and 384-well plates with standard SBS footprints (128.2
mm by 86.0 mm) and heights from 14 mm to 15 mm
Same as base
configuration
Injectors
Base Configuration
Multiple Detectors
Injectors
Can optionally include 2 variable-volume, Watson-Marlow
400F/A pumps with spring loaded tracks for precision, extended
tube life, and easy Santoprene tube loading
Same as base
configuration with total
of 4 injectors
Volume
10 to 100 µL in 1 µL increments
Same as base
configuration
Accuracy
±1 µL all deliveries of requested volume
±1 µL ± 2 %
Precision
±1 µL all deliveries of requested volume
±1 µL ± 2 %
Physical
Base Configuration
Multiple Detectors
Size
22 cm (8.75 inches) high by 42 cm (16.5 inches) wide by 43 cm
(17 inches) deep
57 cm (22.5 inches) wide
for two (2) PMTs with
injectors
Weight
16.4 kg (36 lb)
18.2 kg (40 lb) for two
(2) PMTs with injectors
Line voltage
and
frequency
Input Adapter: 100–240VAC, 2A, 50/60Hz
Output: 24VDC, 6.25A
Additional input adapter
for injector module
Output: 24VDC, 1.5 A
Table A-1: Instrument Specifications