![Molecular Devices SkanWasher 300 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/molecular-devices/skanwasher-300/skanwasher-300_user-manual_1823743065.webp)
SkanWasher 400 User Manual
B-6
Appendix C: Shipment
Prior to transporting the SkanWasher 300, carry out the following steps.
Step 1
Decontaminate the system following
Weekly Maintenance
proce-
dure. If radioactive materials are used, use appropriate decontami-
nate solution for each material.
Step 2
Perform
Empty Head
procedure to empty dispense head of all fluid.
Step 3
Clean exterior surface of instrument with alcohol followed by water.
Step 4
Remove the microplate holder.
Step 5
Secure the wash head using foam and tape.
Step 6
Complete and sign the decontamination form.
Step 7
Ship in the original packaging material and cardboard box.