Maintenance
32
SkanWasher 300 User Guide
Figure 13: Washhead Disconnect
Replacing Probes
Individual probes can be replaced if identified to be the sole problem.
Step 1
Identify if aspirate or dispense probe is to be replaced.
Inlet Probe used for dispense probe (part #1700-0190,
pkg.10/box).
Outlet Probe used for aspirate probe (part #1700-0239,
pkg.10/box).
Step 2
Pull out probe with pliers.
Remove old glue particles from wash head.
Step 3
Insert new probe, burr diameter = 1.10 mm.
Apply Loctite Activator no.7649 on the mounting area on the
probe.
Push probe in until it touches the wash head.
Step 4
Apply glue around probe at the point of entry at the wash head.
Commercially available glue recommended: Loctite
QuickTite Super
Glue Easy Squeeze Gel
or Loctite 290
Loctite Threadlocker Green
(does not set up as quickly as
QuickTite
).
Step 5
Wash excess glue off wash head after 15 minutes using water.
Solenoid valves
Sensor block
Ribbon cable
Air pressure supply
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......