Installation
5
SkanWasher 300 User Guide
Step 4
Connect Waste Reservoir (refer to Figure 1).
Connect the outlet tubing line (red) from the
Waste
port on waste
reservoir to the
Vacuum
port on the side of the instrument (vac-
uum grade tubing required with straight-through quick coupling).
Connect the waste alarm cable from the waste reservoir to the
Waste Alarm
port on the back on the instrument located below
the power receptacle.
Connect vacuum tubing from Vacuum quick connect on waste
reservoir to tubing from vacuum pump. (Hydrophobic filter
should be connected to tubing from vacuum pump with arrow on
filter facing pump.)
Figure 1: Waste reservoir and vacuum pump
Step 5
Connect the Wash and Rinse Liquids (refer to Figure 2).
Connect the pressure tubing (yellow) from the wash buffer and
rinse reservoir to the
Pressure
port on the side of the instrument.
S
CAUTION:
It is recommended
to use the waste collection
vessel provided by Molecular
Devices. These bottles are
implosion proof with a
“tankful” sensor and a built-in
foam reducer.
Waste Reservoir
Vacuum Pump
SW 300
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......